![Molecular Devices SkanWasher 300 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/molecular-devices/skanwasher-300/skanwasher-300_user-manual_1823743011.webp)
SkanWasher 300 User Guide
4
Chapter 2: Installation
Step 1
Verify voltage is properly set for incoming line voltage.
Check that the current voltage setting is pointing to the small
white mark.
Step 2
Change voltage setting, if indicated.
Pry open the fuse compartment with a small screwdriver.
Turn the switch around until the proper setting is pointing to the
small white mark on the bottom of the casing.
Verify correct fuses are in place, change if indicated.
z
230 VAC= 1A fuse
z
115 VAC = 2A fuse
Reinstall switch.
Step 3
Connect the power cord.
Insert the power cord into the receptacle located on the back of
the instrument.
Plug the power cord into the main power outlet.
&
TIP:
Save the packing material for
future use in shipping instrument.
!! IMPORTANT:
Fuse compartment
is located below the main power
receptacle on the back of the
instrument.
S
CAUTION:
Before connecting
the main power, ensure voltage
setting is correct (Step 1).
Содержание SkanWasher 300
Страница 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 30: ......
Страница 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Страница 45: ...Maintenance SkanWasher 300 User Guide 35 Figure 14 Adjustment Procedure Lift Version A 1 4 2 3 B A...
Страница 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Страница 56: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 68: ......