QPix 450 or QPix 460 Colony Picking System User Guide
194
5031498 G
Selecting the Plate and Sanitizing Options
To select the plate and sanitize options:
1. In the Microplates and Sanitise window, from the Source Microplate list, select the source
plate type to hold the colonies to be replicated. If the plate type that you need is not listed,
contact Molecular Devices to add a new plate type to the list. See
For Library Replication, you can select either 96-well or 384-well plates.
For Library Compression, you can select 96-well plates, only.
For Library Expansion, you can select 384-well plates only.
2. Select either to dip the pins a number of times or to stir the wells of the source plate.
To dip the pins a number of times, click Multi Dip and in the Number of Dips field, enter
the number of times to dip the pins.
To stir the wells, click Stir Source.
3. From the Destination Microplate list, select the destination plate type to hold the replicated
colonies. If the plate type that you need is not listed, contact to add a new plate type to the
list.
For Library Replication, you can select only plates with the same number of wells as the
source plate.
For Library Compression, you can select 384-well plates, only.
For Library Expansion, you can select 96-well plates only.
4. Select either to dip the pins a number of times or to stir the wells of the destination plate.
To dip the pins a number of times, click Multi Dip and in the Number of Dips field, enter
the number of times to dip the pins.
To stir the wells, click Stir Source.
5. If you want to make a second copy of the destination plate and have an applicable stacker
lane available, then in the Copies field, enter 2. If you want to clean the pins between
copies of the destination plates, select the Sanitise Between Copies check box.
6. From the Sanitise Profile list, select the Sanitise profile to use for the replication routine.
If the available profiles are not suitable for the replication routine, exit the replication
process and create a new Sanitise profile. See
Creating and Editing Sanitise Profiles on
7. Click Next to set the head and stacker options.
Содержание QPix 450
Страница 1: ...5031498 G August 2022 QPix 450 or QPix 460 Colony Picking System User Guide ...
Страница 6: ...QPix 450 or QPix 460 Colony Picking System User Guide 6 5031498 G ...
Страница 22: ...QPix 450 or QPix 460 Colony Picking System User Guide 22 5031498 G ...
Страница 56: ...QPix 450 or QPix 460 Colony Picking System User Guide 56 5031498 G ...
Страница 102: ...QPix 450 or QPix 460 Colony Picking System User Guide 102 5031498 G ...
Страница 158: ...QPix 450 or QPix 460 Colony Picking System User Guide 158 5031498 G ...
Страница 176: ...QPix 450 or QPix 460 Colony Picking System User Guide 176 5031498 G ...
Страница 218: ...QPix 450 or QPix 460 Colony Picking System User Guide 218 5031498 G ...
Страница 226: ...QPix 450 or QPix 460 Colony Picking System User Guide 226 5031498 G ...
Страница 250: ...QPix 450 or QPix 460 Colony Picking System User Guide 250 5031498 G ...
Страница 256: ...QPix 450 or QPix 460 Colony Picking System User Guide 256 5031498 G ...