Mi Microplate Washer User Guide
48
5032022 C
Defining the Interval Between Each Prime or Rinse Process
Define the number of minutes to wait before running each periodic prime or rinse process.
The defined prime or rinse process repeats after the defined interval of time until the process
is stopped.
Figure 4-22: The Rinse Interval prompt
Button
Action
On the left
Press this button to end the process and return to
Prime/Rinse
. See
Rinsing the Fluidics System on page 31
On the right
The button has no function for this option.
Press this button to increase the interval by 1 minute, up to a maximum of
200 minutes.
Press this button to decrease the interval by 1 minute, down to a minimum of
1 minute.
Press this button to confirm the interval and start the periodic prime or rinse
process. See
Starting the Periodic Prime or Rinse Process on page 49
Table 4-15: Actions of Control Panel Buttons
Содержание MultiWash+
Страница 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Страница 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Страница 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Страница 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Страница 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...