Mi Microplate Washer User Guide
130
5032022 C
Replacing Fuses
If the instrument does not seem to be getting power after switching it on, check to see
whether the supplied power cord is securely plugged into a functioning power outlet and to
the power port on the rear of the instrument.
If the power failed while the instrument was on, check that the power cord is not loose or
disconnected and that power to the power outlet is functioning properly.
If these checks fail to remedy the loss of power, replace the fuses. You can obtain
replacement fuses from Molecular Devices. For fuse specifications, see
.
CAUTION!
Do not touch or loosen screws or parts other than those specifically
designated in the instructions. Doing so could cause misalignment and possibly void
the warranty.
The fuses are located in the fuse carrier on the rear of the instrument.
Item
Description
1
Power switch
2
Fuse carrier
3
Power port
Figure 9-5: Power Switch, Fuses, and Power Port