ImageXpress Pico Personal Imaging System Installation Guide
26
5063666 B
Placing the Instrument
Consider the following when setting up the instrument:
Set the instrument on a stable, dry, flat surface.
Avoid setting up the instrument near any sources of vibration, such as a centrifuge.
During installation, you will need to access the back of the instrument to connect the
cables.
After installation, allow at least 20 cm (7.9 in.) clearance on all sides of the instrument.
Removing the Transport Lock
Remove any tape and the foam transport lock that secures the sample (X-Y) stage the top of
the instrument.
CAUTION!
Retain the shipping box and all packaging materials for future transport
needs.