FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
255
•
Response will be OK confirming the software has attempted to
clear the error.
•
Monitor status for error is cleared.
Loadtips<CR>
Command Description
This command loads the tips onto the pipettor head from position s1.
This command will execute when commanded if tips are not already
loaded and any plate is in the tips position. The instrument is not
capable of confirming that the plate in the tip load position is actually a
tip rack. It is the responsibility of the user to ensure that the plate in
that position is actually a tip rack. Attempting to load tips when a plate,
rather than a tip rack, is in the tip load position will likely result in an
instrument fault and may result in damage to the plate, the instrument,
or both.
Command Parameters
There are no parameters associated with this command.
Example Command Line
loadtips<CR>
Example Response
ok<CR>
Suggested Steps in Command Usage
•
Make sure an experiment is not running.
•
Check the instrument status to make sure tips are not already
loaded.
•
Place tiprack in tipload position via the
loadplate
command.
•
Send
loadtips
command.
•
Response will be OK, confirming that command execution has
begun, or a Condition code related to the command indicating
that execution could not begin.
•
Monitor status for completion, and error status for any other
kind of error.
Command Specific Error Responses
•
Can't load tips, tips on pipettorCondition code is C58
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...