Robotic Integration
246
0112-0109 H
Status<CR>
Command Description
The
status
command returns the status of the instrument. The status
response will be an ASCII text string which includes:
•
Information about the instrument type and firmware version.
•
Whether the instrument is functioning normally or an error
occurred. If an error has occurred, this field will contain an error
code and an error string of less than 300 characters. The error
code and error string will be separated by a comma and space.
•
The instrument status field reports a summary of status
information including whether the instrument is busy, faulted,
error, resetting, aborted, or idle. It also reports whether the
instrument is remote or manual mode. In manual mode, only
status information is available through this interface. All other
commands will not function. Finally, this line reports whether the
software is online (connected to the instrument) or not.
•
Whether an experiment is running or not—do not open or start
another experiment until the first is finished.
•
Name of the plates assigned to each position.
•
Whether plates are present or not on plate locations.
•
Whether plates need to be loaded or removed.
•
Whether the plate handler is busy or idle.
•
Whether the landing pad has a plate or not.
•
Camera temperature. Status returns “not_ok” while camera CCD
temperature is cycling and/or out of range as defined by factory
firmware settings, “off” if camera temperature control is
disabled, or “ok” otherwise. Typically factory setpoint is -60 °C
2 for EMCCD camera or -20 °C
5 for ICCD camera. Also returns
actual CCD temperature and setpoint.
•
Chamber temperature. Status reports “not_ok” while chamber is
not at setpoint temperature (typical factory tolerance of
0.5
°C), “fault” if sensor does not return a reading, “off” if chamber
temperature control has been disabled, or “ok” otherwise.
•
Waste/fill bottle status.
•
Tips on|off.
•
Outer auto door, open/closed—if it is closing it is reported as
closed.
Example Command Line
•
status<CR>
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...