FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
215
Troubleshooting the Yellow Plate
Table 9-6
Troubleshooting the Yellow Plate
Symptom or Error
Message
Possible Causes
Solutions
“Saturation detected,
data may be invalid.”
Incorrect settings.
Set excitation intensity to
80%, exposure length to
0.1 s and for EMCCD gain to
80, for ICCD gate to 3%. If
condition repeats, lower
excitation intensity.
Incorrect LED and/or
emission filter choice.
Check to make sure the
appropriate LED and
emission filters are
selected.
Emission optics are
dirty.
Make sure no dust is on
emission filters.
Bottom of Read plate is
dirty.
Make sure no dust or
fluorescent compounds are
on the bottom of the Read
plate.
Wells are cut off in
Image Viewer.
Plate not properly
placed in Read
position.
Make sure plate is properly
placed and indexed in the
Read position.
Dirty/misaligned
optics.
Contact Technical Support
at
+1-800-635-5577.
Fluorescence counts are
0.
LED power is off.
Increase excitation intensity
of LEDs.
Make sure
Excitation/Emission
Wavelength
in the read
mode does not have
NONE
for the excitation
wavelength.
Camera failure.
Contact Technical Support
at
+1-800-635-5577.
Signal test was not
initiated.
Select
Test Signal
in the
Signal Test dialog box.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...