Running an Experiment
184
0112-0109 H
ion-free buffers. For example, if studying a calcium channel, dye load
cells using a calcium-free buffer and prepare your compound plate
using a calcium containing buffer.
Effect of DMSO on Membrane Potential Assays
High concentrations of DMSO are toxic to many cell lines and affect
compound mixing. Also, DMSO can induce physiological changes in cells
(for example
,
differentiation). If the test compound preparation
contains DMSO, it is important to titrate the DMSO to determine the
maximum concentration that can be used with each indicator cell line
and test compound. With some cell lines that have been tested, there
was no effect on signal level up to 1% DMSO final concentration.
Voltage Sensor Probes Assay Protocol
This section provides the following information and procedures:
•
Materials required for the assay
•
Cell preparation guidelines
•
Dye loading procedure
Required Materials
The following materials are required for running a membrane potential
assay:
Item
Source
FLIPR
®
Tetra System with Voltage
Sensor Probes Optics Kit installed:
FLIPR
®
Tetra System LED Module
390–420 nm FLIPR
®
Tetra System
Emission Filter 440–480 nm FLIPR
®
Tetra System Emission Filter 565–625
nm
Molecular Devices
See Appendix C
Voltage Sensor Probes:
DisBAC2(3) and CC2-DMPE
VABSC-1 Background Suppression Dye
Invitrogen
K1016
K1019
Clear, flat-bottom, black- or clear-wall
384- or 96-well plates
Clear polypropylene source plates
FLIPR
®
System pipette tips, 96 or 384 See Appendix C
Cells in suspension
—
Test compounds
Specific to receptor
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...