Exchanging LED Modules
Uninstalling LED Modules
To uninstall LED banks:
1.
Open the lower front door of the FLIPR
®
Tetra System.
2.
Loosen the two silver
Latching Clamps
securing the excitation
optics module to the left light pipe.
3.
Pull the excitation optics module down and away from light pipe
to disengage the
Power Connectors
on the LED banks from
the light pipe. Be careful not to scratch the bandpass filters
when pulling LED banks out of the instrument.
Remove the
Foam Insert
found on the rear side of the
excitation optics module.
WARNING! During the exchange procedure, you will have
access to precision optical components. Use appropriate care
not to damage them. Specifically, do not touch the surfaces of
the LED filters with your bare fingers or other objects.
Note:
Once the lower door is open, power to the LEDs is
disabled.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...