6 Troubleshooting Procedures
6.6.1 Tilting the Fluidics Module
1 Power
off.
Turn the power off to the instrument. Disconnect the power cords from the instrument
and the cables connected to the computer.
Caution:
Never operate the instrument with the covers removed. Damage to the
instrument can occur. Read and obey all warning labels.
2 Lift up instrument cover.
Press in the handle at the bottom of the top cover and lift the cover back off the
detection module, revealing the fluidics module.
3 Tilt fluidics module.
Using the handle, tilt the fluidics module to the left.
4 Clean
spill.
If you can clean up the spill, or remove the tips, without removing the fluidics module,
do so now. If you need to remove the fluidics module, continue with the instructions in
the section below.
5 Replace the fluidics module.
Using the handle, lower the fluidics module carefully back over the detection module.
Make sure it is firmly seated.
6 Replace cover and reattach cables.
Lower the top cover over the instrument. Reattach the power cord and computer cables.
Emission Fiber
and Restraint
Electrical Connection
to Fluidics Module
Fluidics Module
Excitation Fiber
and Restraint
Tilting the Fluidics Module
116
FlexStation II Operator’s Manual – Rev. D
Содержание FlexStation II
Страница 6: ...FlexStation II Operator s Manual Rev D vi ...
Страница 92: ...4 Applications 86 FlexStation II Operator s Manual Rev D ...
Страница 102: ...5 Service and Maintenance Procedures 96 FlexStation II Operator s Manual Rev D ...
Страница 124: ...6 Troubleshooting Procedures 118 FlexStation II Operator s Manual Rev D ...
Страница 134: ...7 Appendices 128 FlexStation II Operator s Manual Rev D ...