FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
130
0112-0127 B
Problems During Startup
After you start up the SoftMax Pro Software, the Instrument Status bar
should appear as in the illustration on the left, with a temperature
displaying in the temperature box.
Figure 5-1:
Status Bar Detail
If you see a red X in front of the instrument icon, or if no temperature is
displayed in the temperature box (as in the illustration on the right) there
is either a communication problem between the instrument and the
software or the instrument has experienced a fatal error.
1.
Make sure that the FlexStation 3 instrument is plugged in and
turned on.
2.
Make sure that the instrument has finished its warming up
sequence. Wait until the temperature is displayed on the front
panel of the FlexStation 3 instrument.
3.
Click
Edit > Preferences
in the software to view the Preferences
dialog.
Figure 5-2:
Preferences Dialog
Note:
Before starting the SoftMax Pro Software, make sure that the
instrument is powered on.
FlexStation_manual.book Page 130 Tuesday, June 25, 2013 10:22 AM