Digidata 1550 Low-Noise Data Acquisition System User Guide
26
5023399 A
Data throughput problems
•
If decreasing the number of analog input channels or the
sampling rate improves performance, you likely have a data
throughput problem. Verify that the digitizer is connected to a
high-speed USB 2.0 port on your computer. See
Digidata 1550 Digitizer on page 8
. If the digitizer is connected to
a USB hub, remove any other devices connected to the hub, or
bypass the hub.
•
Try the digitizer on a faster computer. CPU speed is only one part
of the equation. Other relevant components include hard disk
speed, RAM speed, and front-side bus speed.
•
To improve data-throughput related acquisition performance
problems in Clampex or AxoScope, try the following:
In
Configure > Lab Book Options
, select
Never log any
events
.
In
Configure > Program Options
, select
Disable screen saver
during data acquisition
.
Contacting Support
Check the Support section of the Molecular Devices website at
www.moleculardevices.com/support
where you will find a link to our
on-line Knowledge Base that contains product-specific questions and
answers to many common issues.
Before you call
To help us to more quickly identify possible problems and known
conflicts, before you call Molecular Devices Technical Support, be ready
to provide the following information:
1.
What is the model and serial number of the digitizer? The serial
number is on a small barcode sticker on the digitizer’s back
panel.
2.
What is the computer environment? To find the system
specifications in Windows 7, right-click
Computer
, and select
Properties
.