Digidata 1550 Low-Noise Data Acquisition System User Guide
20
5023399 A
Step 4
Verify analog inputs.
With AxoScope or Clampex:
1.
Click the
Lab Bench
button (or click
Configure > Lab Bench
)
and select the
Input Signal
tab. For each of the
Digitizer
Channels
(for example,
Analog IN #0
), select a matching
Signal
(for example,
IN 0
), and configure that signal with unity scaling
(for example,
Scale factor
(
V/V): 1
).
2.
Click the
Edit Protocol
button (or click
Acquire > Edit
Protocol
) and then, on the
Inputs
tab, for each of the
Analog IN
Channels, select the matching signal (for example,
IN 0
) from its
list.
3.
Connect a BNC cable from the analog outputs to the analog
inputs to be tested.
4.
Click the
Record
button (or click
Acquire > Record
) to
acquire data. For each analog output signal, use the software’s
Real Time Control
panel to change voltage levels. Either use the
spinners or type a value and then press
Enter
.
5.
Click the
Last Recording
button (or click
File > Last
Recording
) to open the data file. Verify the input signal levels
using the cursors in the window. To display a subset of the
signals, right-click on the data display area, select
Properties
,
and go to the
Show/Hide
tab.