Fluids that spill in the instrument during operation drain through the
prime/soak trough.
Removing and Storing Wash Heads
Heads must always be washed and purged before they are removed. Any
remaining fluid can spill out onto the instrument and potentially cause
damage. Fluid remaining in the head can also lead to the growth of mold
and salt crystals, causing blockage in the dispense and aspiration probes.
CAUTION!
If the trough fluid overfills or spills elsewhere, it leaks
through the bottom of the instrument to the counter beneath and must
be wiped up immediately to prevent damage.
WARNING! Biohazardous Material. Wear protective equipment and
dispose of waste materials in accordance with laboratory site
regulations and procedures. Solutions used in the instrument and
waste materials might be biohazardous.
WARNING! Chemical Waste Disposal Hazard. Wear protective
equipment and dispose of waste materials in accordance with
applicable regulations and procedures. Waste material might contain
toxic, corrosive, and biohazardous material.
Содержание AquaMax 2000
Страница 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Страница 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Страница 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Страница 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Страница 106: ...AquaMax 2000 4000 Microplate Washer User Guide 106 0112 0151K Figure 6 1 Plate stage home sensor Figure 6 2 Plate sensor...
Страница 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Страница 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Страница 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...