Troubleshooting
0112-0151 K
103
Cell Wash Issues
Instrument Error Messages
After performing the recommended tasks in
, touch Stop or
Cancel to continue.
Table 6-8:
Cell Wash Issues
Problem
Possible Causes & Remedies
Coating or cells in wells are
damaged after wash or
aspirate.
Decrease the aspiration rate (96-cell wash
head only).
Decrease the aspiration dwell time
Increase the aspiration probe height.
Coating or cells in wells are
damaged after dispense.
Decrease the dispense rate.
Note:
For the 384-cell wash head, we
recommend to use as high a dispense rate
as possible without damaging the cell
layer.
WARNING! Before manually checking for obstructions, turn off the
power to the AquaMax Microplate Washer.