3. Operating Instructions
Figure 3.7.
Use of the
HORIZONTAL SCALE
knob to improve Vp accuracy.
Figure 3.8.
Final Vp of the cable.
1. Press the
M-FUNC
button until “Smooth” shows up in the upper middle of the screen.
2. Rotate the
M-FUNCTION
knob until the desired level of trace smoothing is obtained.
A comparison of a smoothed (1,024 trace average) TDR trace to an unsmoothed TDR trace at
0
.
7
mρ
(millirho) per vertical division is shown in Figure 3.9.
3.10.5. Sample Resolution
The sample resolution setting establishes the number of sample points per unit time or distance
in the live TDR trace. In general, higher sample resolution sharpens traces at larger horizontal
scales by increasing the effective sampling efficiency of the TDR trace (the ratio of sampled vs
actual change in voltage from sample to sample). For larger horizontal scales it also helps ensure
that all cable faults are sampled adequately enough to be detected and displayed on the
instrument screen. Note that the sample resolution of cable scans can be set separately (see
30
CT100B TDR Cable Analyzers Operator’s Manual