![MOGlabs FSC100 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/moglabs/fsc100/fsc100_manual_1823378006.webp)
2
Chapter 1. Introduction
Error
Error
Frequency
f
Frequency
f
f
0
ERROR OFFSET
∆
f
Figure 1.2:
A theoretical dispersive error signal, proportional to the dif-
ference between a laser frequency and a setpoint frequency. An offset on
the error signal shifts the lock point (right).
Note the distinction between an
error signal
and a
control signal
.
An error signal is a measure of the difference between the actual
and desired laser frequency, which in principle is instantaneous and
noise-free. A control signal is generated from the error signal by a
feedback servo or compensator. The control signal drives an actuator
such as a piezo-electric transducer, the injection current of a laser
diode, or an acousto-optic or electro-optic modulator, such that the
laser frequency returns to the setpoint. Actuators have complica-
ted response functions, with finite phase lags, frequency-dependent
gain, and resonances. A compensator should optimise the control
response to reduce the error to the minimum possible.
The operation of feedback servos is usually described in terms of
the Fourier frequency response; that is, the gain of the feedback
as a function of the frequency of a disturbance. For example, a
common disturbance
f
m
is mains frequency,
f
m
= 50 Hz or 60 Hz.
That disturbance will alter the laser frequency
f
by some amount,
at a rate of 50 or 60 Hz. The effect of the disturbance on the laser
might be small (e.g.
f
=
f
0
±
1 kHz where
f
0
is the undisturbed
laser frequency) or large (
f
=
f
0
±
1 MHz). Regardless, the Fourier
frequency of the disturbance is either 50 or 60 Hz.
Содержание FSC100
Страница 1: ...Fast servo controller FSC100 Version 0 1 2 Rev 2 4 hardware ...
Страница 4: ...ii Contents ...
Страница 14: ...10 Chapter 1 Introduction ...
Страница 22: ...18 Chapter 2 Connections and controls ...
Страница 36: ...32 Appendix B 115 230 V conversion ...
Страница 39: ......