2. Connections and controls
2.1
Front panel controls
The front panel of the
FSC
has a large number of configuration op-
tions that allow the servo behaviour to be tuned and optimised.
Please note that switches and options may vary between hardware
revisions, please consult the manual for your specific device as in-
dicated by the serial number.
INT
SCAN
SCAN+P
LOCK
NESTED
SCAN
LOCK
RATE
BIAS
INPUT
AC
DC
Δ
FAST SIGN
SLOW GAIN
FAST GAIN
DIFF GAIN
GAIN LIMIT
0
10
20
30
60
40
50
MAX
FAST DIFF/FILTER
SPAN
FREQ OFFSET
ERR OFFSET
FAST INT
SLOW INT
EXT
SLOW SIGN
FAST
SLOW
CHA
CHB
FAST ERR
SLOW ERR
RAMP
BIAS
FAST
SLOW
MON1
MON2
Fast Servo Controller
CHB
PD
0
REF
+
–
+
–
100k
200k
OFF
10M
1M
750k
500k
5M 2.5M
50k
25k
100k
250k
OFF
10k
20k
50k
500k
1M
2M
100k 200k
OFF
25
50
75
500
750
1k
100 250
CHA
CHB
FAST ERR
SLOW ERR
RAMP
BIAS
FAST
SLOW
STATUS
0
6
18
24
12
2.1.1 Configuration
INPUT
Selects error signal coupling mode; see figure 3.2.
AC
Fast error signal is
AC
-coupled, slow error is
DC
coupled.
DC
Both fast and slow error signals are
DC
-coupled.
∆
Signals are
DC
-coupled, and the front-panel
ERROR OFFSET
is
applied for control of the lock point.
CHB
Selects input for channel B: photodetector, ground, or a variable 0
to 2.5 V reference set with the adjacent trimpot.
FAST SIGN
Sign of the fast feedback.
SLOW SIGN
Sign of the slow feedback.
5