28
Chapter 3. Operation
below, for wide and narrow spans. These traces were obtained with
an 8 cm long Rb vapour cell at room temperature, using a Zeeman
modulation coil as described in appendix C.
To operate in
AC
locking configuration:
1. Select
AC
locking by setting internal
DIP
switch
7
to
OFF
.
2. Connect the diode,
TEC
and temperature sensor as above. Also
connect a two-channel oscilloscope to the
CHANNEL A, B
and
TRIG
outputs.
3. Connect the photodetector module and optimise the photosig-
nal on
CHANNEL A
. The
MOG
labs
DLC
is designed to operate
best with about 250
µ
W incident on the Si-PIN photodiode.
Lensed and filtered photodiodes are standard, to remove most
background light, and when
AC
locking at 250 kHz modula-
tion frequency, any remaining photocurrent from background
lighting should not be a problem.
BS
PD
250kHz
M
M
BS
Lock-in
ECDL
BS
Servo
Vapour cell + coil
AOM
λ/4
λ/4
f ~ 150mm
f ~ −25mm
Figure 3.5:
Schematic setup for
AC
locking to an atomic transition. PD
is the
DLC
photodetector. BS beamsplitter, M mirror,
λ/
4 retarder. Beam
expanding lenses increase the signal power while minimising saturation
broadening.
Содержание DLC202
Страница 1: ...External Cavity Diode Laser Controller Models DLC 202 DLC 252 DLC 502 Revision 6 00 ...
Страница 4: ...ii ...
Страница 16: ...6 Chapter 1 Introduction ...
Страница 30: ...20 Chapter 2 Connections and controls ...
Страница 42: ...32 Chapter 3 Operation ...
Страница 62: ...52 Appendix C Modulation coils ...
Страница 66: ...56 Appendix D External modulators and injection current modulation ...
Страница 76: ...66 Appendix G Connector pinouts ...
Страница 78: ...68 Appendix H PCB layout ...
Страница 81: ......