Start the cycle by pressing the button corresponding to WASHING CYCLE.
The cycle counter appears in the upper left corner.
WARNING
AT THE END OF THE CYCLE IT IS ESSENTIAL TO VISUALLY INSPECT
THE OBJECTS TO CHECK THEIR PROPER WASHING.
PRE-WASHING CYCLE
The pre-washing cycle is advised to avoid the drying of organic residues on the instruments
and facilitate their removal during the following washing or thermal disinfection cycle.
Rotary instruments, instruments made of untreated aluminium or with decorative anodizing
and instruments with ducts or cavities with reduced opening and high depth, are explicitly
excluded from use.
The reflective surfaces of mouth mirrors can be damaged by ultrasounds.
WARNING
THE LOAD TREATED WITH THIS PROGRAMME IS NEITHER
DISINFECTED NOR WASHED.
USE THIS CYCLE ONLY AS A PREPARATION TO A SUBSEQUENT
THERMAL DISINFECTION OR WASHING PROGRAMME.
AT THE END OF THIS PROGRAMME, HANDLE THE LOAD WITH CARE,
USING THE SUITABLE PERSONAL PROTECTIVE EQUIPMENT.