42
OPERATOR'S MANUAL
EN
WASHING CYCLE (W)
The washing process aims at eliminating organic and inorganic residues present on the
instruments at the end of their use and is suitable for almost all materials, except for the
porous, hygroscopic and water-soluble ones.
Rotary instruments, instruments made of untreated aluminium or with decorative anodizing
and instruments with ducts or cavities with reduced opening and high depth, are explicitly
excluded from use.
The reflective surfaces of mouth mirrors can be damaged by ultrasounds.
WARNING
THE LOAD TREATED WITH THIS PROGRAMME IS NOT DISINFECTED.
USE THE PRE-WASHING PROGRAMME ONLY AS A PREPARATION TO
A SUBSEQUENT THERMAL DISINFECTION PROGRAMME OR IF THE
LOAD TO TREAT CAN NOT BE TREATED AS A MINIMUM AT 80°C OR
FOR MATERIALS NOT CONTAMINATED AND NOT INTENDED TO COME
INTO CONTACT WITH PATIENTS.
AT THE END OF THIS PROGRAMME, HANDLE THE LOAD WITH CARE,
USING THE SUITABLE PERSONAL PROTECTIVE EQUIPMENT.