Chapter 7: Maintaining and Troubleshooting the Arm
Table 20. Troubleshooting - Arm Function
Problem
Cause
Solution
Arm moving without
command
IMU zero position
changed
1.
Put the arm into Standby Mode
2.
Check and secure the IMUs on your
feet.
3.
Put your feet flat on the ground
4.
Take the arm out of Standby Mode
Sweat near EMGs
1.
Put the arm into Standby Mode
2.
Wipe the sweat from the EMG
electrode and skin with a dry cloth
3.
Take the arm out of Standby Mode
Ham radios or walkie
talkies are affecting the
arm
1.
Put the arm into Standby Mode
2.
Move the arm at least 0.5 m (20 inches)
away from any ham radios or walkie
talkies
3.
Take the arm out of Standby Mode
Arm not moving
Arm is in Standby Mode
Put the arm into Hand Mode
Arm is off
1.
Shake the IMUs to wake them
2.
Power the arm on
Arm is faulted
Table 19, Troubleshooting - Wrist
Input device not
connected
1.
Power the arm off
2.
Check and tighten all connections at the
ACI
3.
Power the arm on
Cannot change Modes
Input device not
connected
1.
Power the arm off
2.
Check and tighten all connections at the
ACI
3.
Power the arm on
Input device not working
Contact Technical Support. See
Cannot change grips
Hand is not fully open
1.
Put the arm into Hand Mode
2.
Fully open the hand
3.
Change grips
Input device not working
Contact Technical Support. See