System Two LED Installation and Service Manual (1003385 Rev. C)
Page 54
Section 5
Installation and Removal for the Sterilizable Handle Assembly (Camera Lights only)
Before the light can be used, the Sterilizable Handle Assembly must be installed.
The D2C comes equipped with an interchangeable Sterilizable Handle Assembly. The available interchangeable
options are the Sterilizable Handle Assembly and a HD Camera Assembly.
To install the Sterilizable Handle Assembly it is first necessary to remove the plastic Sterilizable Handle, then turn
the mounted light head so that the LED’s face vertically towards the ceiling. Position the Handle Assembly so that
the alignment pins align with the corresponding alignment holes in the light head. Using one hand to prevent the
light head from moving gently push the Handle Assembly until it is firmly inserted into the light head. Then fasten
the three captive screws and reattach the plastic Sterilizable Handle. The light head is now ready to use.
To remove the Sterilizable Handle Assembly, reverse the above sequence.
Figure 32: Attaching the Sterilizable Handle