971 UniMag
™
Operation manual
19
To ensure that no unauthorized personal are able to change transducer setup and parameters, the
transducer lock function can prevent direct access to parameter changes. Transducers delivered with pre-
configured custom specified parameters (special part number) are per default locked and will reply with
“NAK180”, if the user tries to change locked parameters. The unlock procedure must be executed to change
these parameters.
Disable lock function command:
Command:
@253FD!UNLOCK;FF
Command reply:
@253ACKFD;FF
Enable lock function command:
Command:
@253FD!LOCK;FF
Command reply:
@253ACKFD;FF
Standard transducer (7 digits part number: 971-xxxx)
Factory default:
Transducer unlocked
Special configuration transducer (11 digits part number: 971-xxxx-xxxx)
Factory default:
Transducer locked
If the transducer is delivered with special configuration, the lock function will only be
temporally disabled and will be enabled again after cycling power cycle or executing
the enable lock command.
The 971 transducer can be delivered with factory locked tamperproof settings for
safety interlock applications. This option is defined in the special settings. If
delivered with factory lock the transducer settings can only by changed by return of
gauge to MKS.
The User Switch command defines if the Cold Cathode sensor voltage is turned on by level trig or pulse trig.
For details see Cold Cathode sensor control.
Command:
@253SW!OFF;FF
Command values:
ON,OFF
Command reply:
@253ACKOFF;FF
Query:
@253SW?;FF
Query reply:
@253ACKON;FF
Factory default:
OFF
The transducer test command can be used to visually indentify a transducer. If the test mode is enabled the
LED will flash with a 1 sec. cycle.
Command:
@253TST!ON;FF
Command values:
ON,OFF
Command reply:
@253ACKON;FF
Query:
@253TST?;FF
Query reply:
@253ACKON;FF
Factory default:
OFF
Содержание UniMag 971
Страница 1: ...P N 100017131 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...
Страница 36: ...971 UniMag Operation manual 33 mm Inch mm Inch...
Страница 37: ...971 UniMag Operation manual 34 mm Inch...
Страница 40: ...971 UniMag Operation manual 37...
Страница 41: ...971 UniMag Operation manual 38...
Страница 42: ...971 UniMag Operation manual 39...
Страница 44: ...P N 100017133 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...