MKS UniMag 971 Скачать руководство пользователя страница 20

 

971 UniMag

 

 

Operation manual

 

 

17 

Cold Cathode protect setpoint 

The Cold Cathode protect setpoint automatically turns off the Cold Cathode high voltage if the cold cathode 
measurement exceeds 5.00E-3 Torr for 120 seconds. The protect setpoint can be turned off (disabled) or 
time set between 0 and 999 seconds. 
 
 

Command:  

@253PRO!ON;FF 

 

Command values: 

ON (120), OFF, 0 to 999 seconds 

 
 

Command reply: 

@253ACK120;FF 

 

Query:   

@253PRO?;FF 

 

Query reply: 

@253ACK120;FF 

 

Factory default: 

OFF  

 

 

If FP is set to “ALWAYSON” the protect setpoint is automatically disabled. (See page 15) 

The 971 is factory calibrated when delivered and in many applications further calibration is not required.  
If the sensor element has been contaminated or damaged by process gases, adjustment of zero and full 
scale can be executed to compensate for small measurement errors. 
 

 

Pressure unit calibration 

The transducer can provide digital and analog output in torr, mbar and Pascal pressure units. 
When changing pressure unit all parameters such as setpoint settings are automatically converted to the 
new unit, so it will represent the same pressure level. All pressure parameters such as setpoint settings and 
calibration values must be entered in the actual transducer unit setting. 

 

Change of pressure unit calibration setup: 
 
 

Command:  

@253U!PASCAL;FF 

 

Command values: 

TORR, MBAR, PASCAL 

 
 

Command reply: 

@253ACKPASCAL;FF 

 

Query:   

@253U?;FF 

 

Query reply: 

@253ACKTORR;FF 

 

Factory default: 

TORR 

 

 

The torr unit is most common in the US and mbar is most common in Europe. Pascal is the official pressure 
unit as specified by SI (from the French 

Le 

S

ystème 

I

nternational d'Unités

) and is widely used in Asia. 

 

  

Содержание UniMag 971

Страница 1: ...P N 100017131 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...

Страница 2: ...ed and approved in the handling and cleaning of contaminated materials before the product will be accepted by HPS TM for repair and or replacement In the course of implementing this policy HPS TM Cust...

Страница 3: ...l 10 Setpoint relays 12 Calibration and adjustment 17 Factory default 18 Transducer lock function 19 User Switch Command 19 Transducer test 19 Status Query Commands 20 Analog output 21 Query Command l...

Страница 4: ...MKS Denmark ApS Ndr Strandvej 119G DK 3150 Hellebaek Denmark Tel 45 44 92 92 99 Fax 45 44 92 94 99 E mail mksdenmark mksinst com MKS Instruments HPS Products 5330 Sterling Dr Boulder CO 80301 USA Tel...

Страница 5: ...can cause permanent damage to the transducer or interference to measuring performance Refer to Electrical connections description page 7 Fuse The 971 power supply input has an internal thermal fuse T...

Страница 6: ...and sensor sealing type Transducers can be delivered with customer configuration of various parameters like setpoint settings and these specials have additional 4 digits after the regular part number...

Страница 7: ...ons are accelerated towards the anode and will ionize molecules by collision The magnetic field deflects the electrons causing them to spiral as they move across the magnetic field to the anode This s...

Страница 8: ...s delivered with customer specified parameters the User Switch is disabled For enabling the switch see page 19 The red green LED status indicator has the following stages 1 GREEN Normal operation 2 2...

Страница 9: ...temperature gradients and direct cooling like air condition air stream or direct heating like a pump exhaust stream Bake out The transducer electronics can withstand 85 o C 185 o F when the power is t...

Страница 10: ...e analog out and negative analog out pins to a differential input voltmeter or an analog to digital A D converter Do not connect the negative side of the analog output to the negative side of the powe...

Страница 11: ...the transducer 971 I O Connector P N 971 x1090 0001 6 pin Hirschmann PIN Description 1 Identification resistor 5 1K 2 Analog Output 3 Analog Output 4 Power 9 30 VDC 5 Power return 6 Chassis P N 971 x...

Страница 12: ...quality and communication speed but relative long cable spans up to 1 200m 4 000 ft is possible There are 2 wires other than ground that are used to transmit the digital RS485 signal The 971 uses half...

Страница 13: ...e the user enters into the command syntax Each query or command must begin with the attention character and end with the termination FF Command syntax for an information query device address query FF...

Страница 14: ...nd all transducers will transmit a reply Commands sent with address 255 will be executed by all transducers on the network but the transducers will not transmit replies For example use address 254 to...

Страница 15: ...izing spike protection Inadequate protection can cause permanent damage to the transducer or interfere with the analog output signal Always ensure that inductive inrush currents do not exceed relay co...

Страница 16: ...6 00E 6 FF Reply 253ACK6 00E 6 FF 4 Enable setpoint OFF ON Command 253EN1 ON FF Reply 253ACKON FF Setpoint setup by PDR900 Controller 1 Edit Setpoint Setpoint 1 Setpoint Value 1 Enter set point value...

Страница 17: ...KON FF Factory default ON The 971 with integrated display provides real time pressure readout pressure trend and setpoint status The display setup can be changed via the intuitive menu interface The s...

Страница 18: ...er can either be momentarily pulse trigged or low level trigged Analog Level trig factory default When enabling level trig the cold cathode sensor will be on when pin 13 is shorten to ground pin 4 and...

Страница 19: ...ion The digital pressure output can provide 3 digit or 4 digit values however the resolution is limited in certain parts of the measuring range 1 00E 7 to 1 00E 8 Torr 2 digits resolution 1 50E 8 5 00...

Страница 20: ...process gases adjustment of zero and full scale can be executed to compensate for small measurement errors Pressure unit calibration The transducer can provide digital and analog output in torr mbar a...

Страница 21: ...point 3 enable EN1 OFF Setpoint safety delay SPD ON Cold Cathode protect setpoint PRO OFF Switch enable SW OFF Analog out 1 AO1 30 1 1 If the transducer is delivered with other analog output than stan...

Страница 22: ...transducer is delivered with special configuration the lock function will only be temporally disabled and will be enabled again after cycling power cycle or executing the enable lock command The 971...

Страница 23: ...ry reply 253ACK971 FF Part Number PN Specifies transducer part number Query 253PN FF Query reply 253ACK971 11030 FF Serial Number SN Specifies transducer serial number Query 253SN FF Query reply 253AC...

Страница 24: ...d from Torr to Pascal or mbar the analog output scaling will change as well so it represents 0 5 VDC decade Torr or 0 5 VDC decade mbar or Pascal Torr mbar Vout Torr mbar Vout 1 0E 8 1 5000 1 0E 5 3 0...

Страница 25: ...g Change of analog output setup Command 253AO1 15 FF Command values 10 to 319 xy First digit x 1 PR1 Cold cathode pressure value assignment Second digit y 0 MKS Standard 0 5 VDC decade 1 Edwards APG L...

Страница 26: ...2 3 16 7 00 7 50E 1 1 00 100 8 00 2 37 31 6 3 160 9 00 750 0 1 000 100 000 10 00 Analog output calibration 5 Inficon MPG400 Pfeiffer PKR251 emulation PTorr 10 Vout 11 46 1 667 Vout log10 PTorr 0 6 6 8...

Страница 27: ...Edwards AIM S SL emulation provides a strongly non linear output Torr mbar Pascal Vout 1 00E 8 1 33E 8 1 33E 6 2 5 1 80E 8 2 40E 8 2 40E 6 2 5 4 40E 8 5 87E 8 5 87E 6 3 6 10E 8 8 13E 8 8 13E 6 3 2 8 3...

Страница 28: ...67E 4 6 67E 2 8 656 1 00E 3 1 33E 3 1 33E 1 9 000 5 00E 3 6 67E 3 6 67E 1 9 799 Analog out calibration 18 Pfeiffer IKR251 The Pfeiffer AIM X XL emulation provides a log linear output Torr mbar Pascal...

Страница 29: ...nt 1 3 hystereses switch value xxxEN1 FF xxxEN2 FF xxxEN3 FF xxxACKOFF FF Set point 1 3 enable status OFF PIR MicroPirani or CC Cold Cathode xxxSD1 FF xxxSD2 FF xxxSD3 FF xxxACKBELOW FF Set point rela...

Страница 30: ...tion xxxBR 19200 FF xxxACK19200 FF Set communication Baud rate 4800 9600 19200 38400 57600 115200 230400 xxxAD 123 FF xxxACK123 FF Set transducer communication address 001 to 253 xxxRSD OFF FF xxxACKO...

Страница 31: ...pressure cycles can the transducer withstand A Cold Cathode sensor element is very mechanical robust to pressure changes and there are no limits on the number of pressure cycles In applications where...

Страница 32: ...sensor chamber can be cleaned using solvents compatible with 304 stainless steel 403 stainless steel Ceramic Al2O3 Viton and epoxy resin Q Can the sensor element be changed if contaminated A Yes the c...

Страница 33: ...been executed Incorrect pressure value at high pressure Contaminated sensor Transducer repair required Other gas or gas trace present than transducer gas setting Cold Cathode does not provide measure...

Страница 34: ...anode module can be changed from the flanges side without disassembling of the transducer The anode module p n 970 ANODEKIT is available as spare part The basic accuracy of the cold cathode sensor ca...

Страница 35: ...nce 100 m max Setpoint relay contact endurance 30VDC 1A load 100 000 min Setpoint relay contact endurance 30VDC 0 2A load 2 000 000 min Setpoint relay response time 100 ms Materials exposed to vacuum...

Страница 36: ...971 UniMag Operation manual 33 mm Inch mm Inch...

Страница 37: ...971 UniMag Operation manual 34 mm Inch...

Страница 38: ...25ft RS485 100013674 10 m 33ft RS485 PDR900 Connectors cables Part number Description 100010757 Setpoint Relay 3 pin connector 100013638 Analog output 8 pin connector 100013686 Analog output cable 3 m...

Страница 39: ...disturbances characteristic of industrial scientific and medical RF equipment Class B Group1 Immunity EN 61000 4 2 Electrostatic discharge EN 61000 4 3 Radiated RF electromagnetic fields EN 61000 4 4...

Страница 40: ...971 UniMag Operation manual 37...

Страница 41: ...971 UniMag Operation manual 38...

Страница 42: ...971 UniMag Operation manual 39...

Страница 43: ...l volume 32 LED Status Indicator 5 Lock function 19 Magnetic interference 2 Manufacturer 20 Model 20 NAK Code 10 NAK180 13 19 30 Operating temperature 32 Part number 3 Part Number 20 PDR900 35 PDR900...

Страница 44: ...P N 100017133 971 UniMag Transducer Operation and Installation Manual Revision C January 2010...

Отзывы: