4.10 Convectron Gauge Analog Output Signal
If the Convectron Gauge capability is installed, a voltage output
signal proportional to the common logarithm of the pressure
indication is provided on the rear panel of the Convectron Gauge
module via a standard 1/8 in. miniature phone jack.
If graphed on log-linear axes, the output voltage is linear with respect
to the log of pressure. The analog output is 1 volt per decade of
pressure with a factory adjusted output of 0 volts at 1.0 x 10
-4
Torr.
Offset adjustments are provided on the top edge of the Convectron
Gauge module that allow shifting the voltage corresponding to 1 x
10
-4
Torr between -7 volts and +1 volt.
The voltage signal is smooth and continuous throughout all the decades of pressure measurement.
This format is useful for computerized data acquisition because a simple equation (finding the
common antilogarithm) may be programmed to calculate pressure from the voltage output.
The equation is: P
i
= 10
V-4
Torr/mbar, or P
i
= 10
V-2
Pascal
Figure 4-17
Top View of Convectron
Gauge Module Showing
Analog Output Offset
Adjustment
Figure 4-18
Convectron Gauge Analog Output vs. Pressure
where
Pi = pressure indication,
V = analog output voltage
Gauge analog output (V)
-7
-6
-5
3
4
5
6
7
8
9
10
1
10
10
3
10
2
10
-4
10
-3
10
-2
10
-1
10
2
10
5
10
4
10
3
10
-2
10
-1
1
10
lower adjust limit
factory adjustment
2
1
0
-1
-2
-3
-4
upper adjust limit
CONVECTRON
Pressur
e (Torr)
Pressur
e (Pa)
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......