2
Initial Setup Procedures
2-6
Series 370 Stabil-Ion Instruction Manual - 370119
Process Control setpoints 1 (PC1) and 2 (PC2) are controlled by the ionization gauge or gauges.
Setpoint 1 operates relay 1. Setpoint 2 operates relay 2. Relays 1 and 2 are also shown as K1 and K2
on the printed circuit board for reference during IG assignment. IG1, IG2, or both can be assigned
to operate either setpoint 1 or 2. Use Table 2-2 to assign the IG’s to the setpoints.
2.1.3.2 Relay Polarity Setting
The relays can be set to activate as pressure either
rises above or falls below the setpoint. A DIP switch is
provided for each setpoint. Refer to the numbers on
the switch for the setpoint number. Use Table 2-3 to
assign the relay polarity settings.
The switches are factory preset as shown on Table 2-3
for relay activation below the pressure setpoint. This is
most commonly desired when you want the relay to
be de-energized under a high pressure condition.
Table 2-2
IG Process Control Channel Switch Settings
Switch
Settings
Function
Switch
Settings
Function
Setpoint 1 can turn ON when IG1
is ON or when IG2 is ON.
(factory setting)
Setpoint 2 can turn ON when IG1
is ON or when IG2 is ON.
(factory setting)
Setpoint 1 can turn ON when IG1
is ON and will turn OFF when
IG2 is ON.
Setpoint 2 can turn ON when IG1
is ON and will turn OFF when
IG2 is ON.
Setpoint 1 will turn OFF when
IG1 is ON and can turn ON when
IG2 is ON.
Setpoint 2 will turn OFF when
IG1 is ON and can turn ON when
IG2 is ON.
Setpoint 1 will always be OFF.
Setpoint 2 will always be OFF.
Figure 2-9
Partial Top View of Process
Control Module
9
10
OFF
OFF
7
8
OFF
OFF
9
10
OFF
ON
7
8
OFF
ON
9
10
ON
OFF
7
8
ON
OFF
9
10
ON
ON
7
8
ON
ON
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......