Initial Setup Procedures
2-5
Series 370 Stabil-Ion Instruction Manual - 370119
2.1.3.1 Ion Gauge Assignment for Process Control
A maximum of 6 process control setpoints are available
depending on the particular option you have specified. A
setpoint is defined as the combination of vacuum gauge
indication setpoint circuitry and the associated relay which is
actuated when the pressure indication corresponds to the
setpoint.
NOTE 1:
Setpoint 1 is user-
assignable to IG1 or
IG2, or IG1 and IG2.
Setpoint 2 is user-
assignable to IG1 or
IG2, or IG1 and IG2
independent of how
Setpoint 1 is assigned.
Figure 2-6
Two Channel Process Control - Cat. No. 370108
NOTE 2:
Setpoints 3 and 4 are
factory assigned to
Convectron Gauge A.
Setpoints 5 and 6 are
factory assigned to
Convectron Gauge B.
Figure 2-7
Six Channel Process Control - Cat. No. 370107
Figure 2-8
Partial Top View of Process Control
Module
T
W P
PROCESS CONTROL
1
2
3
4
5
6
SET
Rear Panel
Connector Pin
Designators
NO C NC
Note 1
2 independent sets
of relay contacts
Setpoint 1
O
IG1
O
DEGAS
O
POWER
O
IG2
Normally Open
Common
Normally Closed
Setpoint 2
D
H
A
IG
2.13 -8
A
B
5.67 -2
745
1
2
3
4
5
6
T
W P
D
H
A
E
J
B
R
M
U
PROCESS CONTROL
1
2
3
4
5
6
SET
Rear Panel
Connector Pin
Designators
NO C NC
"
"
Note 1
6 independent sets
of relay contacts
Setpoint 1
2
3
5
V
S
X
F
C
K
"
"
"
4
6
Note 2
O
IG1
O
DEGAS
O
POWER
O
IG2
IG
2.13 -8
A
B
5.67 -2
745
1
2
3
4
5
6
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......