1
System Components
1-5
Series 370 Stabil-Ion Instruction Manual - 370119
Figure 1-7
Stabil-Ion Power Supply
1. Line power fuse
6. Line voltage selector switch
2. Grounding lug (connect to earth ground with
#12 AWG conductor)
7. Connector for interconnect cable to Control Unit
3. Degas power fuse
8. Connector for Stabil-Ion Gauge 1
4. Line voltage selector switch
9. Connector for Stabil-Ion Gauge 2
5. Line voltage selector switch
10. Connector for power cable
9
!
!
ALL 3 SWITCHES MUST BE SET TO THE SAME AC LINE VOLTAGE
POWER 50-60 HZ 220 WATTS
IG2
IG1
DEGAS
FUSE
CONTROL UNIT
POWER
FUSE
10
1
2
3
4
5
6
7
FOR AC LINE VOLTAGE BETWEEN 90V TO 130V SET ALL SWITCHES TO 115V,
USE 3A, F 250V POWER FUSE, AND USE 2A, F, 250V DEGAS FUSE
FOR AC LINE VOLTAGE BETWEEN 180V TO 240V SET ALL SWITCHES TO 230V,
USE 1.6A, F, 250V POWER FUSE, AND USE 1A, F, 250V DEGAS FUSE
8