All items in the kit may not be needed, depending on the mounting configuration.
Contact an MKS Applications Engineer for solutions to other mounting configurations.
10.5 AC Power Cord
The 946 Controller has a standard female IEC 60320 connector for connecting to a 100-240 VAC, 50/60
Hz power source. Use only a harmonized, detachable cord set with conductors having a cross-sectional
area equal to or greater than 0.75 mm
2
. The power cord should be approved by a qualified agency such
as UL, VDE, Semko, or SEV.
Properly ground the controller and vacuum system.
The 946 Controller is grounded through the ground conductor of the power cord. If the protective ground
connection is lost, all accessible conductive parts may pose a risk of electrical shock. Plug the cord into a
properly grounded outlet only.
Do not exceed the manufacturer's specifications when applying voltage. Electrical
shock may result.