902B Abs Piezo
Operation manual
23
Analog output
The 902B transducer provides a voltage output as function of pressure. The 902B is available with a
selection of factory configured outputs determined by the part numbers listed below. The default analog
output calibration is “14” corresponding to the Linear10 functions stated below for the units: torr, mbar and
pascal.
Fla
ng
e
In
te
rf
ac
e
An
al
og
o
ut
Co
nn
ec
to
r
Re
la
ys
En
cl
os
ur
e
se
ali
ng
902B-
1
1
0
2
0
-
0 0-10 V, (Linear 10 mV/Torr/mbar)
1 0-5 V, (Linear 5 mV/Torr/mbar)
2 2-10 V (LOG10)
3 1-5 V (LOG5)
4 1 V/decade
5 0-10 V (100 mV/Torr)
6 1-9.8 V
Analog output calibration = 0 (Linear10 10mV/Torr/mbar)
P
Torr
= V
out
× 100
V
out
= P
Torr
/ 100
P
mbar
= V
out
× 100
V
out
= P
mbar
/ 100
P
Pascal
= V
out
× 10,000
V
out
= P
Pascal
/ 10,000
Torr
mbar
Pascal
V
out
1.0
1.33
1.33E+2
0.010
5.0
6.66
66.66E+2
0.050
10.0
13.3
1.333E+3
0.100
50.0
66.66
6.66E+3
0.500
100.0
133.3
1.333E+4
1.000
500.0
666.6
6.666E+4
5.000
1,000
1333.2 1.3332E+5
10.000