Chapter 3
56
Micro-Ion
®
Module Instruction Manual - 356007-GP
You may also ground pin 11 on the 15-pin subminiature D connector to
lock the CAL, Gauge, Degas, and Trip Points buttons on the module.
To unlock the buttons, remove the pin 11 ground.
If you’ve locked the buttons by grounding pin 11, you cannot use the KBU
command via RS–485 to unlock them. Remove the ground before sending
the KBU command.
KBU – Unlock the
Keyboard
After you’ve locked the CAL (or ATM), Gauge, Degas, and Trip Points
buttons, use the keyboard unlock (KBU) command to unlock them.
KBU command from host:
#01KBU
↵
Response from module:
*01 PROGM OK
↵
If you’ve locked the buttons by grounding pin 11 on the 15-pin
subminiature D connector, the KBU command will not change the locked
state. Remove the ground before sending the KBU command.
FAC – Reset the Values to
Factory Defaults
Table 3-19 lists default settings for the RS-485 version of the module. After
you’ve reconfigured the module, you must cycle power to the module
before the new settings will take effect.
You may restore parameters listed in Table 3-19 to their default values by
sending a factory reset (FAC) command.
FAC command from host:
#01FAC
↵
Response from module:
*01 PROGM OK
↵
Reset command from host:
#01RST
↵
Response from module:
None
Table 3-19 Factory Default Settings for RS-485 Module
Parameter
Default Setting
Digital communication
• Parity: 8 data bits, no parity, 1 stop bit
• Baud rate: 19.2 kbaud
• Module address: 01
• Address offset: 00
Emission current threshold
• 5 x 10
–6
Torr (7 x 10
–6
mbar, 7 x 10
–4
pascal) with decreasing pressure
• 1 x 10
–5
Torr (1 x 10
–5
mbar, 1 x 10
–3
pascal) with increasing pressure
Relays
• Trip point: Sets to 0.0 vacuum pressure; relay must be configured to be operable
• Hysteresis and activation direction: Relay activates at 10% above vacuum pressure trip
point (–) with decreasing pressure
Pressure output units and
optional pressure display
Pressure unit is Torr, mbar, or pascal as specified on sales order and catalog number tag
Содержание 356 Micro-Ion Plus Series
Страница 2: ......
Страница 24: ...24 Micro Ion Module Instruction Manual 356007 GP...
Страница 58: ...Chapter 3 58 Micro Ion Module Instruction Manual 356007 GP...
Страница 68: ...Appendix A 68 Micro Ion Module Instruction Manual 356007 GP Figure A 1 Automatic Filament Switching Flow Chart...
Страница 70: ...70 Micro Ion Module Instruction Manual 356007 GP Figure A 3 Manual Filament Switching Flow Chart...
Страница 74: ...Index 74 Micro Ion Module Instruction Manual 356007 GP...
Страница 75: ......