Electromagnetic Flow Meter
FW: 841012/842009
Wiring Schematic #2
for the "Remote Display and Multiple Converters Wiring" example on the
opposite page (page
Local flow sensor 1
2-wire twisted shield cable:
Connect lead "1" to slot "A",
lead "2" to slot "B" and the
twisted shield to slot "-".
Local flow sensor 2
2-wire twisted shield cable:
Connect lead "1" to slot "A",
lead "2" to slot "B" and
twisted shield to slot "-".
4-wire twisted shield cable:
Connect lead "1" to slot "A",
lead "2" to slot "B",
lead "3" to slot "+" and
lead "4" and twisted shield to slot "-".
Remote display unit
4-wire twisted shield cable:
Connect lead "1" to slot "A",
lead "2" to slot "B",
lead "3" to slot "+" and
lead "4" and twisted shield to slot "-".
1 2 3 4 5 6 7 8 9
A
B
207920
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
4 and
shield
1 2 Shield
2
1
3
Содержание MagFlux
Страница 6: ...6 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 12: ...12 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 38: ...38 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 106: ...106 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 144: ...144 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 149: ...Index 5 GB MagFlux Manual 0812 Electromagnetic Flow Meter Electromagnetic Flow Meter This page intentionally left blank...