![MJK MagFlux Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/mjk/magflux/magflux_installation-and-user-manual_1812930030.webp)
30
GB MagFlux Manual 0812
Electromagnetic Flow Meter
See also application examples in Appendix G. Digital Input/Output Con-
nections on page
Digital Input
MagFlux
®
has one digital input which is activated with a voltage higher
than 10 V DC and de-activated with a voltage lower than 5 V DC.
Digital Input
Terminal
Designation
DI
Com
DI
Max. 30 V DC
The digital input (DI) can be programmed for the following functions:
-
start and pause batch 1 and batch 2 counters
-
reset counters R Totalizer Forward and Reverse
-
reset counter R Totalizer Sum
DI shares the common terminal (Com) with DO 1.
See also application examples in Appendix G. Digital Input/Output Con-
nections on page
Local (compact) Flow Sensor
Connect the compact (local) flow sensor to the flow converter with the
wires coming from the flow sensor as shown below.
Important:
Terminals 1 and 4 must be shunted!
Compact Flow Sensor
Terminal
Signal
Designation
Colour from sensor
1
Liquid GND
Built-in liquid
ground electrode
Black/shield
2
E1
Electrode
Red
3
E2
Electrode
White
4
GND
Ground
Black
5
L1
Coil
Blue
6
L2
Coil
Brown
7, 8 and 9
not used
Содержание MagFlux
Страница 6: ...6 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 12: ...12 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 38: ...38 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 106: ...106 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 144: ...144 GB MagFlux Manual 0812 Electromagnetic Flow Meter This page intentionally left blank...
Страница 149: ...Index 5 GB MagFlux Manual 0812 Electromagnetic Flow Meter Electromagnetic Flow Meter This page intentionally left blank...