![Mizuho VACB-3006B Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/mizuho/vacb-3006b/vacb-3006b_operators-manual_1812837054.webp)
50
8
Before
contacting
your
distributor
for repair
Trouble shooting
The following situations can occur even if the operating table is not out of order.
Check the following points before contacting your distributor for repair.
Problem
Possible cause
Measures
The table cannot
be switched on.
1. The connector of the control unit is not
connected properly.
1. Insert the connector completely.
(
→
P21)
A function button
on the control unit
does not function.
1. The connector of the control unit is not
connected properly.
2. E button is not pressed simultaneously
with the function button.
3. Protection circuit is activated.
1. Insert the connector completely.
(
→
P21)
2. Press E button and functional switch
simultaneously.
(
→
P27 – 34)
3. Wait for about 7 minutes to operate.
(
→
P15, P21)
The brake button
on the control unit
does not function.
1. The emergency brake release handle is in
a release (UNLOCK) position.
2. After the brake was unlocked with the
emergency brake release handle, the brake
was not unlocked by the control unit.
1. Turn the emergency brake release
handle toward “LOCK”.
(
→
P28)
2. Unlock the brake by using the control
unit. (
→
P28)
The table cannot
be moved.
1. The auxiliary brake is secured.
1.Unlock the auxiliary brake.
(
→
P29)
The table does not
slide.
1. The lock mechanism of sliding the table is
temporarily deactivated.
1. Perform the reset operation.
( → P36 )
In case that the condition is not improved even if the above measures are taken, please contact
your distributor or us for repair.
Содержание VACB-3006B
Страница 2: ......
Страница 4: ......
Страница 5: ...1 Introduction 1 Introduction ...
Страница 6: ...1 Introduction ...
Страница 8: ...1 Introduction ...
Страница 9: ...2 Before use 2 Before use ...
Страница 17: ...3 Component Identification 3 Component Identification ...
Страница 43: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Страница 46: ...42 5 Maintenance and inspection ...
Страница 47: ...6 Specifications 6 Specifications ...
Страница 50: ...46 6 Specifications ...
Страница 51: ...7 When the control unit can not be used 7 When the control unit can not be used ...
Страница 53: ...8 Before contacting your distributor for repair 8 Before contacting your distributor for repair ...
Страница 56: ...52 8 Before contacting your distributor for repair ...
Страница 57: ...Appendix ...