17
4
Operation
Charging the battery
CAUTION
•
Connect the product to the power source provided
with the protective grounding to prevent the risk of
an electrical shock.
•
Make sure to use the dedicated power cord with
the "MIZUHO" logo.
•
Before inserting the power cord into the power
connector, check that the power connector does
not have any fluid in it nor is dusty.
•
When turning off the power completely, disconnect
the power cord from the medical grade outlet.
•
If the battery deteriorates, it will not be available for
the operating table when AC power is not supplied
due to power outage etc.
1.
Connect the power connector of the product and the
medical grade outlet with the power cord.
POWER CORD
BATTERY INDICATOR
POWER SWITCH
FOOT SWITCH
E
F
Power switch
Medical grade outlet
Power connector
Power cord
2.
When turn on the power switch, battery charging starts.
While charging, the battery indicator sequentially flashes orange.
3.
When the battery indicator sequentially flashes orange and
green, charging is completed.
POWER CORD
POWER SWITCH
FOOT SWITCH
BATTERY INDICATOR
E
F
E
F
E
F
E
F
BATTERY INDICATOR
Red
Orange
Orange Green
Charging in progress
(sequentially flashes
orange)
Charging complete
(sequentially flashes
orange and green)
Needs charging
(lit red)
NOTE
•
Make sure to charge the
battery when initially using
the purchased product, or
when it has not been used
for a long time. The battery
naturally discharges itself
when it is not being used
and is being stored.
•
If the battery lamp on the
control unit blinks or the
battery indicator on the lower
part of the operating table
lights up only red while using
the operating table with the
battery power, charge the
battery immediately. When the
battery is discharged, only the
AC power is available and do
not use the operating table
with the battery power.
•
The operating table battery life-
span is about 2 years. Once it
reaches its life-span, request
your distributor or Mizuho for a
battery replacement.
•
The life-span for the battery
varies greatly depending on
operating conditions. The
battery could degrade quicker
if charging and discharging the
battery are repeated frequently
after using the operating table
for short operations.
It is recommended that you
charge the battery once a
week on weekends, since
it takes 17 hours to fully
charge the battery.
•
If the battery is discharged
soon even after charging, the
battery may be degraded.
Request repairs from your
distributor or Mizuho.
•
While the battery is being
charged, you can operate
the operating table by using
on the control unit.