Before contacting for repairs
8
. Before contacting for repairs
Checking causes and countermeasures
The following problems can occur even if the operating table is not malfunctioning. Check the following points
before requesting repairs.
Status
Possible cause
Measures
The table cannot be
turned on.
The connector of the control unit is not
connected properly.
Insert the connector completely.
(
à
The battery is fully discharged.
Charge the battery.
(
à
A function switch on
the control unit does
not function.
The connector of the control unit is not
connected properly.
Insert the connector completely.
(
à
E switch is not pressed simultaneously
with the function switch.
Press E switch and functional switch
simultaneously. (
à
Protection circuit is activated.
Wait for about 20 minutes to operate.
(
à
The brake switch on
the control unit does
not function.
The emergency brake release handle is
in a release (UNLOCK) position.
Turn the emergency brake release
handle toward “LOCK”. (
à
After releasing the brake by the
emergency brake release handle, the
brake has not been unlocked by the
control unit.
Unlock the brake by the control unit.
(
à