44
7
T
roubleshooting
7
. Troubleshooting
When the control unit cannot be used
The extra switch should be used only in an emergency.
CAUTION
•
Unlike the control unit, the extra switch has no
function to halt the operation of the buttons.
•
Always watch movement of the operating table
when you operate the extra switch.
•
If any parts come in contact with each other,
immediately stop the operation. Otherwise, the
operating table may get damaged.
z
Use the extra switch to operate the operating table
1.
Open the lid of the extra switch.
Control panel
2.
Press any function button on the control panel according
to the desired direction.
Reverse
Trendelenburg:
Head up
Lateral tilt:
Right down
Back plate
bending: Up Lift: Up
Trendelenburg:
Head down
Lateral tilt:
Left down
Back plate
bending: Down
Lift:Down
Control panel
(enlarged view)
NOTE
The operating table moves
while the switch is being
pressed. The table stops
once the maximum angle is
achieved.