MES-CK08-089-00EN-0 

PREPARED:

 

(2022-01-12) (Version 5) 

Instruction for Use 

 

Trade Name: Lateral Cassette Holder 

 

Contraindication/Prohibition 
This product is only validated for use with Mizuho operating tables. 

Symbol mark for labeling 

  MD 

  : Medical Device

 

Specifications 
1.

 

Shape 

 

 
 

 

 
 

 

 
 
 

 

 

The Rail clamp shown above is Clamp for Flat Bar Ver.2 

Code No. 

Product Description 

Compo

nents 

Weights 

(kg) 

08-089-03 

Lateral  Cassette  Holder  with 
Clamp 

①④⑤⑩ 

*1 

1.4 

08-089-03-NC 

Lateral Cassette Holder 

①⑩

 

*2 

1.2 

08-089-05 

Lateral  Cassette  Holder  with 
Clamp Ver.2 

①②③⑩ 

*3 

1.6 

08-111-00 

Clamp for Flat Bar 

④⑤ 

*4 

0.2 

08-111-11 

Clamp for Flat Bar Ver.2 

②③ 

*5 

0.4 

08-117-02 

Clamp for Flat Bar (R1) 

⑤⑦ 

*6 

0.2 

08-117-10 

Clamp for Flat Bar Ver.2 (R1) 

③⑥ 

*7 

0.4 

08-117-12 

Clamp for Flat Bar (R2) 

⑤⑨ 

*8 

0.2 

08-117-13 

Clamp for Flat Bar Ver.2 (R2) 

③⑧ 

*9 

0.4 

Note: The number corresponds to those used in 1. Shape. 
*1: *1 comes with *4. 
*2: Please select *6, *7, *8 or *9 to use *2. 
*3: *3 comes with *5. 
*4 to *9: The size of mountable side rail is different. 

2.

 

Material 

 Lateral cassette holder: Stainless steel rod 

②⑥⑧

 Clamp for Flat Bar Ver.2: Stainless steel 

 Fixing handle: Aluminum alloy, Stainless steel rod 

④⑦⑨

 Clamp for Flat Bar: Stainless steel 

 Fixing handle: Plastic molding, Stainless steel rod 

 Fixing knob: Plastic molding, Stainless steel rod 

Note: The number corresponds to those used in 1. Shape. 

Intended purpose 

This is an accessory to an operating table. 
It is used for mounting an X-ray cassette to have an X-ray for a lateral 
part of a patient. 

Intended user 

This product is to be used by health care professionals, including but not 
limited to surgeons, nurses and biomedical technicians. 

Instructions for use 

 

 
 
 
 
 
 
 
 
 
 

The Rail clamp shown above is Clamp for Flat Bar Ver.2 

 

1.

 

Attach the Rail clamp to the Side rail of an operating table. 

2.

 

Insert the Lateral cassette holder into the Rail clamp and tighten the 
Fixing handle to temporarily secure the Lateral cassette holder to 
the Side rail.

 

3.

 

Attach the Cassette to the Lateral cassette holder and tighten the 
Fixing knob to fix the Cassette storage frame.

 

4.

 

Move  and  adjust  this  product  to  a  position  appropriate  for  the 
patient’s posture.  

5.

 

After the adjustment, tighten the Fixing handle to secure the Rail 
clamp and the Lateral cassette holder. 
Note:  The following size of cassettes can be set with the sliding 

function of the Cassette storage frame. 
254 mm x 305 mm 
203 mm x 254 mm 

Warning/Caution 
1.

 

Warning 
Ensure  all  fixing  handles  are  tight.  Loose  condition  can  cause  the 
product to come off or move, which may result in a patient injury. 

2.

 

Important caution 

2-1. When working with an operating table, take care not to allow this 

product to make contact or interfere with the table top or with other 
tools  and  appliances  used  in  combination  with  this  product. 
Otherwise the product could break. 

2-2. For hygiene, be sure to use sterilized drapes on the areas on this 

product where the patient comes into contact with it. 

Storage/Life 
1.

 

Do not store the device in high temperatures or in areas with high 
humidity where the temperature has drastic variations. 

2.

 

Service life of this product: 7 years 
(Subject to the specified maintenance and inspection and is stored 
properly.) 

Be sure to read this instruction manual before using this product. 
Keep this manual available for reference when needed.  

Cassette frame 

Note: Cassette is NOT included 

Side rail 

Fixing handle 

Cassette frame 

Clamp for Flat Bar 

 

Cassette 
storage 
frame  
(sliding 
function)

 

Lateral cassette 
holder 

Fixing knob

 

Cassette 

Cassette storage 
frame 

Rail clamp 