MES
-
CK08-075-11EN-0
PREPARED: (2022-10-18) (Version 3)
Instruction for Use
Trade Name: Base Unit Attachment
1/2
Be sure to read this instruction manual before using this product.
Keep this manual available for reference when needed.
Warning
When using operating table accessories, always pay close
attention to the patient. Patients left in the same posture for an
extended period of time are at risk for developing nerve paralysis
and/or bedsores.
Contraindication/Prohibition
1. This product is only validated for use with Mizuho operating
tables.
2. Use only appliances suitable for the intended use of this
product. Incorrect use could cause breakage of this product as
well as that of tools and appliances used in combination with
it.
Symbol mark for labeling
MD : Medical Device
Specifications
1.
Shape
①
Frame
④
Insertion hole of the Attachment
②Insertion shaft
⑤Fixing handle of the Attachment
③Fixing handle
Code No.
Product Description
08-075-11
Attachment for Mayfield Head Support
2.
Weight
2.1 kg
3.
Material
①
Frame: Stainless steel
②
Insertion shaft: Stainless steel
③
Fixing handle: Synthetic resin
④
Insertion hole of the Attachment: Stainless steel
⑤
Fixing handle of the Attachment: Synthetic resin
Note: The numbers correspond to those used in 1. Shape.
Intended purpose
This is an accessory for an operating table. This product is used to
attach the base unit such as the Head Supporter.
Intended user
This product is to be used by health care professionals, including
but not limited to surgeons, nurses and biomedical technicians.
Instructions for use
1. Remove the head plate from the operating table.
2. Attach the Base Unit Attachment to the back plate insertion
holes of the operating table.
3. Tighten the fixing handles of the back plate.
4. Attach the base unit to the insertion holes of the Attachment.
5. Tighten the fixing handles of the Attachment.
6. Attach the component parts to the base unit according to
the instructions for use of the base unit.
Note: Refer to the instruction manual of the base unit for more
usage information of the base unit.
Warning/Caution
1.
Warning
Ensure all fixing handles are tight. Loose condition can cause
the products to come off or move, which may result in a patient
injury.
2.
Important caution
2-1.When working with an operating table, take care not to allow
this product to make contact or interfere with the table top
or with other tools and appliances used in combination with
this product.
2-2.For hygiene, be sure to use sterilized drapes on the areas on
this product where the patient comes into contact with it.
Storage/Life
1.
Do not store the device in high temperatures or in areas with
high humidity where the temperature has drastic variations.
2.
Service life of this product: 7 years
(Subject to the specified maintenance and inspection and is
stored properly.)
Unit
:
mm
Fixing handle
Fixing handle
Fixing handle of
the Attachment
Base Unit
Insertion hole of
the Attachment
Insertion hole
Insertion hole
Operating table back plate
Base Unit Attachment