Foreword 

To obtain  the  highest  performance  and  the  longest service  life from  your 
Linear  Gage,  carefully  read  this  manual  thoroughly  prior  to  setup  and 
operation. After reading this manual keep it near the Linear Gage for quick 
reference.  The  specifications  of  this  gage  and  the  description  in  this 
manual are subject to change without prior notification. 

 

Safety Precautions 

To  ensure  operator  safety,  use  the  instrument  in  conformance  with  the 
directions and specifications given in this User’s Manual. 

 

Precautions for Use 

Carefully avoid the following attempts to protect the instrument from failure 
and malfunction. 

 

IMPORTANT 

 Do not apply sudden shocks including a drop or excessive force to the 

linear gage. 

 Do not disassemble or modify the gage. 

 Do  not  use  and  store  the  gage  at  sites  where  it  is  exposed  to  direct 

sunlight or at extremely hot or cold sites. 

 To use the gage highly accurately, avoid sites where the temperature 

will change abruptly. 

 Absolutely  do  not  apply  an  electric  engraver  to  the  gage.  The  high 

voltage may damage electronic parts. Also, do not use the gage at sites 
where it is subject to large electronic noises. 

 Do not exert load on the spindle in the perpendicular direction and do 

not twist the spindle. 

 Do  not  clamp  the  stem  too  tightly,  since  the  spindle  will  not  move 

smoothly. 

 Do not apply excessive tension to the cable or do not bend it forcibly. 

 To perform stable measurement, allow at least 10 minutes after turning 

on the power. 

 Do not set up the origin point at either end of the stroke. 

 If the gage is used in combination with other instruments, the maximum 

performance  could  not  be  obtained  depending  on  environmental  and 
operating conditions. Take those conditions into consideration prior to 
use. 

 The  functions  and  performance  will  not  be  guaranteed,  if  the  gage  is 

used in other conditions than those specified. 

 Take  sufficient  damage-preventive  processing  (safety  measures), 

should this gage have been at fault. 

 

Caution

 

Please be careful enough when handling  the knife edge or blade 
type contact point, since there is a possibility of injury at exchange 
and use. 
 

Electromagnetic Compatibility (EMC) 

This product complies with the EMC Directive. Note that in environments 
where electromagnetic interference exceeds EMC requirements defined in 
this  directive,  appropriate  countermeasures  are  required  to  assure  the 
product performance. 

 

1. Name and Dimension of Each Part 

 
 
 
 
 
 
 
 
 
 
 
 

 

2. Mounting the Gage 

(1)To mount the gage on another instrument or a fixture, clamp the 

φ

8 stem. 

(2)It is recommended to use a slotted holder or a split bushing for the mount 
structure.

Recommended tightening torque in Example 1

4.0 to 5.0kgf

cm

 

 

 

 

 

 

 

 

 

 

 

 

 
 

 

IMPORTANT 

 Absolutely avoid pressing the stem directly 

with  set screws.  (The  built-in bearing may 
be damaged.) 

 Mount  the  gage  so  that  the  spindle  is 

directed  perpendicular  to  the  measured 
surface. If the gage is mounted at an angle 
to the measured surface, an error may be 
generated in measurement results. 

 

  3. Precautions in Protecting the Gage from Dust and Water 

 

IMPORTANT 

 The output connector plug (counter side) is not protective structured. Install 

the gage at a place where it is not splashed directly with water or oil. 

 If the cable covering is broken, liquid will penetrate into the gage inside 

due  to  capillary  phenomenon.  This  will  cause  damage  to  the  gage. 
Immediately repair the cable. 

 Be  greatly careful  not  to  damage the  rubber  boot  due  to chips,  etc. If 

the rubber boot is damaged, dust-proof and water-proof function will be 
deteriorated. Immediately replace or repair the rubber boot. 

  The materials including rubber which are used for the rubber boot and 

other  sealing  parts  are  not  universal  against  diversified  coolants  and 
chemicals.  If  those  parts  deteriorate  unusually,  consult  the  nearest 
Mitutoyo Service Center. 

 Each  part  of  the  gage  is  sealed  up,  and  therefore  must  not  be 

disassembled. If any part is disassembled, the rated performance will 
not be obtained. Do not absolutely disassemble the gage. 

 

4. Specifications of Gage Output Signals 

1) 

Output connector

RM12BPE-6PH (Hirose) 

2) 

Pin assignment 

 

 

 

 

 

 

3)

 

Output signal level

 

   

φA、φA、φB、φB(

TTL line driver AM26LS31 or equivalent

) 

4)

  Input connector and recommended input circuit 

Suitable connector

RM12BRD-6S (Hirose) 

 

 

 

 

 

 

 

Stem 

 

Rubber boot 

 

Spindle 

 

Contact point 

 

Signal cable 

 

Output connector 

unit:mm 

M9.5×0.5 

30 

φ

8 

   

0 

-0.009 

 

Thrust stem 

mounting screw 

Ring of discriminating 

  a model 

Cable length  2m 

φ

12 

 o

m

or

e(

st

ro

ke

) 

 

⑤ 

RM12BPE-6PH 

(Hirose) 

① 

② 

③ 

⑥ 

④ 

14

 

6.

 

 

 

 

cross section 

cross section 

 

② 

① 

① 

Notch position 
(90°with respect to tap) 

φ

8 

unit:mm 

M4×0.7 thru 

・Example 2. 

・Example 1. 

M4×0.7 thru 

4.5 drill,  7.5 countersink, 
4.4 deep 

10 

20 

7.5 

+0.020 
+0.005 

 

7.5 

15 

 

 

15

 

15

 

1 

16 

φ

10 

7.5 

15 

+0.020 
+0.005 

 

②Split bushing 

1 

15

 

φ

8 

+0.020 
+0.005 

 

φ

10 

    0 

-0.015 

 

 

Error=L0-L1 

θ

 

 

Power supply to the gage head 

Power voltage:5V(4.8V to 5.2V) 
Ripple voltage:200mVp-p or less 
Current consumption:120mA max. 

 

 

Linear Gage LGK series 

 

 

No.99MBC081B② 
SeriesNo.542 

 

Pin No. Signal name 

 

+5V 

 

φA 

 

φB 

 

φA 

 

GND 

 

φB 

 

※ 

 

1 

2 

3 

4 

6 

5 

 

 

φ

A,

φ

B 

φ

A,

φ

B 

Recommended input circuit 

LGK  output circuit 

26LS32  or equivalent 

33pF 

51

Ω

 

GND 

+5V 

GND 

+5V 

 

 

 

 

 

 

 