1. Names of Components
②
①
③
④
⑤
⑩
⑨
⑥
⑧
⑦
⑪
⑫
CGF-X
①
Clamp screw
②
Indicator holder
③
Grip
④
Pipe (outer sleeve)
⑤
Measuring head
⑥
Supporting nut
⑦
Anvil
⑧
Adjustment washer
⑨
Guide
⑩
Contact point (contact rod)
⑪
Guide stopper screw
⑫
Contact rod holder
Tips
No guide (guide plate) is provided with CG-S10A.
2. Preparation for Measurement
1) Setup
NOTICE
• Confirm that the measuring head or screw is not loosened before use.
• Do not insert, remove, or turn the indicator with the clamp screw tightened, as it will lead to
breakage.
• If the insertion hole of the indicator holder is accidentally deformed, correct it by inserting a
ø8 mm rod for the metric type and a ø9.53 mm rod for the inch type.
• Secure fixation may be impossible when the indicator or the indicator holder insertion hole or
the clamp screw is dirty. Clean them in advance.
1
3
Dial protection cover
Clamp screw
1
Loosen the clamp screw and insert the indicator into the indicator holder.
2
Tighten the clamp screw to fix the indicator.
Tips
Firm tightening is possible by turning with an object such as a coin inserted in the groove on the
clamp screw head.
3
When required, attach a dial protection cover (optional depending on models).
4
For models with a measurement range of 160 mm or more, the direction of the guide unit must
be switched.
1. Push the guide until it hits the end lightly.
2. Turn 90 degrees clockwise.
3. Let go.
⇨
The guide returns to the reference point position, and the product is ready for measurement.
1
2
Safety Precautions
To ensure operator safety, use this product in conformance with the directions, functions
and specifications given in this User’s Manual.
Use under other conditions may compromise safety.
NOTICE
• Do not disassemble or modify. This may cause damage.
• Do not use or store the product in a place with sudden temperature changes. Adapt the
product to room temperature before use.
• Do not store the product in a place with high humidity or a lot of dust.
• Do not apply impact or excessive force to this product.
• Be sure to perform reference point setting before measurement.
• Remove dust, cutting chips, etc. before and after use.
• After use, clean and perform anti-rust treatment for the body, anvils, adjustment washers,
etc. Insufficient cleaning may cause accuracy or operation to suffer.
• For periodic calibration or precision measurement, wear thick gloves in order to reduce
changes in the indicated value caused by the transmission of body temperature.
• Indicators with rubber bellows such as waterproof dial gages cannot be used.
Tips
The bore gage is a comparator. It will not function as a measuring instrument if used alone.
An indicator such as a dial gage and a reference gage such as a set ring or micrometer are
required to use the product.
Bore Gage
User's Manual
No.99MAJ005A
CG-SA
CG-AX
CGB-X
CGF-X
CGM-X
Model/Code No. List
Model
Code No.
CG-S10A
511-209
CG-S18A
511-201
CG-S.4"A
511-214
CG-S.7"A
511-205
CGB-35X
511-761
CGB-60X
511-762
CGB-150X
511-763
CGB-160X
511-764
CGB-1.4"X
511-781
CGB-2.5"X
511-782
Model
Code No.
CGB-6"X
511-783
CGB-6.5"X
511-784
CG-35AX
511-701
CG-60AX
511-702
CG-150AX
511-703
CG-160AX
511-704
CG-250AX
511-705
CG-400AX
511-706
CG-1.4"AX
511-731
CG-2.5"AX
511-732
Model
Code No.
CG-6"AX
511-733
CG-6.5"AX
511-734
CG-10"AX
511-735
CG-16"AX
511-736
CGF-35X
511-415
CGF-60X
511-416
CGF-150X
511-417
CGF-1.4"X
511-418
CGF-2.4"X
511-419
CGF-6"X
511-420
Model
Code No.
CGM-100X
511-803
CGM-160X
511-804
CGM-250X
511-805
CGM-400X
511-806
CGM-600X
511-807
CGM-800X
511-808
CGM-4"X
511-833
CGM-6.4"X
511-834
CGM-10"X
511-835
CGM-16"X
511-836
Model
Code No.
CGM-24"X
511-837
CGM-32"X
511-838
en