Printed in Japan
Date of publication: November 1, 2017
Mitutoyo Corporation 20-1, Sakado 1-Chome, Takatsu-ku, Kawasaki-shi, Kanagawa 213-8533, Japan
1. Part Names
1
2
3
4
5
6
7
8
9
11
12
13
10
1 Gage head
2 Main body unit
3 Contact point
4 Attachment
5 Lever
6 Output connector
7 Battery holder (with protection cap)
8 Display unit (LCD)
9 Control panel part
10 DATA/HOLD switch
11 Battery ON/OFF switch
12 SET switch
13 MODE in/mm switch
(in/mm switch is for in/mm product only)
14
15
16
24
17
23
18
22
19
21
20
14 Hold display
15 Sign display
16 Function lock display
17 Power supply voltage warning display
18 INC display
19 Tolerance judgment display (-NG)
20 Tolerance judgment display (OK)
21 Tolerance judgment display (+NG)
22 P.SET display
23 Unit display (in is for in/mm product only)
24 Measured value display (tolerance judgment
enlarged display)
2. Setting the Battery
Note
• Be sure to use SR44 (silver oxide battery).
• Be careful not to damage battery terminals when installing the battery.
• If the battery holder is not correctly mounted, it may cause abnormal display or failure.
• Do not use a sharp-pointed tool to remove the battery holder and do not pry out the battery holder, to prevent
damage to it.
• Replacing the battery clears all settings. Reset them.
• When disposing the battery, comply with ordinances and regulations.
At purchase, the battery is not set in this product. Follow the procedure below to set the battery.
1
Use a driver or other tool to remove the battery holder.
2
When replacing a battery, remove the old battery from the battery holder.
3
Set a new battery (SR44) in the battery holder as shown in the figure below.
4
Mount the battery holder.
>> "- - - - - -" lights.
5
Press the SET switch twice.
>> Absolute measurement (ABS) mode starts.
If it does not enter the absolute measurement (ABS) mode, remove the battery once, and then set it again.
1
4
5
2
3
3. Powering On/Off
To turn on the power, press the battery ON/OFF switch.
To turn off the power, hold the battery ON/OFF switch (for
two seconds or longer).
Note
• This product starts up in the last used measurement mode upon power-on.
(For measurement modes, see "4. Switching Measurement Modes".)
• If this product is not turned on even when the battery ON/OFF switch is pressed, the battery may be exhausted.
Replace it.
• Power-off during any setting cancels the setting and restores the previous status.
• When the power is turned on, a value other than the actual one or "Err 30" may be displayed. This is not a failure. It
will return to the actual value by operating the lever.
4. Switching Measurement Modes
This product has the following two measurement modes:
• Absolute measurement (ABS) mode: Sets the reference point with a ring gage (reference gage) and measures a
diameter of the workpiece.
• Incremental measurement (INC) mode: Zero-sets the display value with a reference master and measures the
difference in diameter between the reference master
and the workpiece.
To switch the measurement modes, hold the SET switch
(for two seconds or longer).
Note
When the SET switch is pressed in the incremental measurement (INC) mode, the display value is zero-set.
Safety Precautions
To ensure operator safety, use this product in conformance with the directions and specifications given in this
User's Manual.
Use under other conditions may compromise safety.
• Always keep batteries out of reach of children, and if swallowed, consult a physician immediately.
• Batteries should never be short-circuited, disassembled, deformed or come in contact with extreme
heat or flames.
•
If battery alkaline liquid comes in contact with the eyes, flush eyes immediately with clean
water and consult a physician. If battery alkaline liquid comes in contact with the skin, flush
the exposed area thoroughly with clean water.
• Never attempt to charge the primary battery or reverse the positive-negative terminals when
mounting. Improper battery handling or mounting may cause the battery to explode, cause
battery leakage and/ or serious bodily injury or malfunctioning.
• Always handle the sharp measuring faces of this product with care to avoid injury.
Note
• Do not disassemble or modify this product. It can cause failure.
• Avoid using or storing this product where there is significant temperature change. Prior to use, thermally
stabilize the product at room temperature.
• Do not store this product in a high-humidity or dusty environment. Do not use this product where it could be
splashed with coolant, etc.
• Do not apply sudden shocks (such as dropping) or excessive force to this product.
• Be sure to perform reference point adjustment before measurement.
• Oil or cutting chips adhering to or rust on the sliding part of the contact point results in incorrect operations.
Wipe off oil and cutting chips after use.
• To clean this product, use a soft cloth soaked in a diluted neutral detergent. Do not use any organic solvent
(thinner, etc.). It may deform or damage this product.
• Do not use an electric engraver to put numbers or marks on the product. It can cause failure.
• The supplied battery is used to check functions and performance. It might not provide the specified life.
• If the product will not be used for more than three months, remove the battery from and store it properly.
Otherwise, liquid may leak from the battery and damage the product.
• The warranty shall not apply if the product falls or is damaged as a result of fair wear and tear, including
battery drain.
• Do not twist this product or let it hang while it is fitted in a workpiece.
• Only use the supplied contact points. Do not remove the contact point.
• The measured value differs depending on whether the entire surface of the contact point is used for
measurement or only at the tip because of the mechanism of this product. Make the reference point adjustment
under the same conditions as for the measurement.
• Be sure to make the reference point adjustment before measurement or after detach/attach of the gage head
and extension rod (option).
WARNING
CAUTION
BOREMATIC
User's Manual
No.99MAB035A
SR44
SET
ON/OFF
ON/OFF
SET
SET
=
> 2 sec
=
< 1 sec
en
(Press twice)
Incremental measurement (INC) mode
Absolute measurement (ABS) mode
Icons of Switch Operations