5.9 Changing Resolution
The
gauge
has
selectable
display
resolution, which is 0.1mm
and 0.01mm.
When the resolution is set to 0.01mm, the tested piece surface should be smooth to get
accurate test result. When measuring rough surface or coarse grained materials, it is suggested to
use low resolution.
5.10 Memory Management
5.10.1 Store a Record
By simply pressing the
key after a new measurement reading appears, the measured
thickness value will be saved to current selected data group. It is added as the last record of the
group.
5.10.2 View the Saved Record
This function provides the user with the ability to view the records in a desired data group
previously saved in memory. Following is the steps:
Activate the Memory
Manager dialog as right
figure.
Press
to move the
cursor; Press
or F3 key
to open the View Record
Data dialog, see next
figure.
Press
to move the
cursor
to
the
desired
record.
Press F3 to delete the
focused record.
Press F2 to clear all the
records of this group.
Press
/F1 to exit.
5.10.3 Select As Current Data Group
There are 100 data groups (F00 ~ F99) inside the gauge that can be used to store the
measurement values. At most 100 records (thickness values) can be stored to each group. You
can change the destination data group to store the measured values as following.