MIT GAL-2L-GAL-2H Скачать руководство пользователя страница 1

 

 

G O L D I L U X 

L I G H T 

M E T E R S

 

 

USER MANUAL 

 

GAL - 2L, GAL - 2H,  

GAL - 3L with GALP probes 

GENERAL DESCRIPTION 
EXTERNAL FEATURES AND CONTROLS 
OPERATING INSTRUCTIONS

 

MAINTENANCE AND PRECAUTIONS

  

CALIBRATION

 

SPECIFICATIONS

 

WARRANTY INFORMATION 

GENERAL DESCRIPTION 

The GOLDILUX series of light meters are designed to 
measure illuminance of the visible spectral radiation in the 
units of lux or optionally in footcandles. 
An external probe may be plugged into a light meter. In this 
case the detector in the light meter is automatically 
disconnected and the displayed reading corresponds to the 
quantity and the units measured with the 

probe

 
The liquid crystal display of the metres (GAL-2L, GAL-2H) is 
auto ranging over one decade. External probes (e.g. GALP-
1L) have a built-in 

probe gain selector switch

 with two 

different gains and thus cover a wider measuring range than 
the meters with built-in detector. The probe gain settings are 
multiplied as indicated on the probe labels and the switch 
setting indicated by the 

probe range factor

. Probes 

combined with a non-specific meter (display - GAL-3L) offer 
the best combination with the widest possible measuring 
range. 

 

EXTERNAL FEATURES AND CONTROLS

 

Fig. 3 indicates the layout and function of the switches and 
connectors and other features available on the light metres. 

 

 

The ON/OFF switch (A in Fig. 3) is located on the left side 
of the meter housing. To switch ON move it forward. The 
meter is OFF when the switch is in the lower position. 
 
The HOLD button (B in Fig. 3) can be used to "freeze" the 
display for convenient reading by the operator. This hold 
button operation can be achieved for up to 5m distance 
from the instrument by means of a 

remote hold cable 

(optional extra)

 

plugged into connector C in Fig. 3. A simple 

contact closure between the two 

mono plug

 connector pins 

activates the hold ("freeze") function. 
 
The ANALOG OUTPUT of the meter is D in Fig. 3. It can be 
used to supply a 2V output for a full-scale reading. It has an 
output impedance of 10 k  and can be used to record light 
levels on a chart recorder or any other suitable data 
acquisition system. A 

stereo plug

 should be used to 

connect the equipment to the analogy output socket. 
 

OPERATING  INSTRUCTIONS

 

1.  Remove the instrument from its case and mount or 

place it in the desired measurement position. 

 
 

 

2.  Switch on the instrument and check its zero with the 

cap firmly on the detector. Should it be necessary to 

adjust the meter reading to zero (e.g. for calibration 

purposes) adjust it via the potentiometer accessible 

through the latch opening in the open battery 

compartment.  

3. Remove 

the 

protective 

cap from the detector being 

used. 

4.  Take readings. Check on the units (Kilolux or lux or 

footcandles as indicated by the label). 

 

N.B.: A 9V battery is necessary for operating the instrument 

and this has to be inserted in the battery compartment at 

the back of the meter as shown in Fig 3. 

 

MAINTENANCE AND PRECAUTIONS

 

When not in use, always put the protective cap on the 

detector and keep the instrument in a safe place. 

There is a warning "battery low" on the LCD display when 

the 9V battery needs to be replaced. 

If necessary, clean the detector with a soft clean cloth or 

tissue moistened with alcohol.  Dry and polish lightly with a 

dry tissue. 

 
 

3

 

CALIBRATION

 

1) Meters 
a)  Meters with a built-in detector (e.g. GAL-2L): adjust the 

meter reading to zero via the potentiometer accessible 
through the opening in the open battery compartment. 
and the cap firmly on the detector, 

b)  Expose meter to a known illuminance, emitted by a 

light source of the type for which the calibration is 
desired. Alternatively, produce a stable illuminance with 
a suitable light source of the desired type and measure 
it with a calibrated light meter. 

c)  Adjust the meter with the calibration adjustment (F in 

Fig. 3) until it reads correctly. 

d)  Seal the hole (F in Fig. 3) with a suitable calibration 

sticker. 

 
2) Probes 
a)  Plug the probe into the meter. Set the probe’s range 

selector switch (A in Fig 4) to the lowest setting. 

b)  With the dustcap firmly in place (D in Fig 4) check the  

meter reading of zero. The potentiometer P5 indicated 
on the rear label of the probe has been disabled.   

 
 

 
Return instrument for repair if zero reading cannot be 
obtained. 

c)  Expose probe as in 1 b) above. 
d)  If required the reading is adjusted by turning the 

potentiometer P1 as indicated on the rear label of the 
probe, until the reading is correct.  

e)  Set the probe range factor to the highest range factor 

and adjust P2 (as marked on the rear label) until a 
reading of exactly 10 times lower is achieved than with 
the selector switch in the lowest range factor setting. 

f)  For probes with integral dose option (e.g.GALP-2L) - 

please refer to the manufacturer. This range has been 
discontinued in favour of a microprocessor range of 
instrumentation.   

g)  To make the adjustments on the potentiometers P1 and 

P2 it is necessary to pierce the rear label of the probe. 
Seal again with suitable plastic stickers after the 
calibration is completed.  

 
NOTE : 

Calibrations should only be performed by 
trained metrologists in a recognized calibration 
laboratory. 

 
 

Отзывы: