Страница 3: ...rotected by copyright Distribution and copying in any form of the content is prohibited without the explicit and written consent of Miris AB name ID revision issue date article number name article number standard Miris Ultrasonic Processor User Manual M S003 22a 2 2018 04 03 36 06 201 Miris Ultrasonic Processor 08 04 201 2014 30 EU Electromagnetic Compatibility Directive 2014 35 EU Low Voltage Dir...

Страница 4: ...WARRANTIES AND DUTIES 5 SAFETY INFORMATION 6 MIRIS ULTRASONIC PROCESSOR USER GUIDE 7 Chapter 1 INSTALLATION 7 Chapter 2 SETTINGS 9 Chapter 3 OPERATING THE MIRIS ULTRASONIC PROCESSOR 10 Chapter 4 CLEANING ROUTINES 12 Chapter 5 TROUBLESHOOTING AND PROBLEM SOLVING 14 ABOUT MIRIS ULTRASONIC PROCESSOR 15 TECHNICAL SPECIFICATIONS 15 WORKING PRINCIPLES OF THE INSTRUMENT 15 ...

Страница 5: ...sor will be used according to the instructions given in the user manual The Miris Ultrasonic Processor will not be altered without written approval of MIRIS AB MIRIS will be notified immediately if any injury occurs in any association with the Miris Ultrasonic Processor and will be allowed prompt and thorough examination of the Miris Ultrasonic Processor in question MIRIS will not be held responsi...

Страница 6: ...High voltage is present in the power supply Do not remove the cover Service can only be performed by an authorized service technician Ask your local distributor or contact Miris Never operate the power supply unless it is connected to the converter Never secure anything to the probe Never touch a vibrating probe Never allow a probe to vibrate in air for more than 10 seconds Always use input voltag...

Страница 7: ...iris Ultrasonic Processor User Manual INSTALLATION 7 MIRIS ULTRASONIC PROCESSOR USER GUIDE Chapter 1 INSTALLATION 1 Sound reducing cabinet 2 Display and keypad Figure 1 The Miris Ultrasonic Processor 1 2 ...

Страница 8: ...ts check the label on the back of the unit Probe processing capacity For efficient homogenisation it is important to not exceed the processing capacity of the probe The 3 mm diameter probe is for milk volumes 3 10 ml 3 Height adjustment knob 4 Convertor housing 5 Probe 6 Display with current settings 7 Probe output power indicator 8 Digit input 1 9 9 Clear 10 Enter Review 11 Start Stop 12 On 13 Of...

Страница 9: ... pressing Timer Use the key pad to enter sample processing time given as hours minutes seconds in the format H MM SS Press Enter to confirm The set time is displayed PULSE Pulsating mode Pulse should be turned off Press Pulse Press Clear to erase current setting Press Enter to confirm Pulse is now turned off shown on the display as AMPLITUDE Amplitude should be set to 75 Press Ampl Use the key pad...

Страница 10: ...at the Miris Ultrasonic Processor probe is clean and that the tip is evenly polished Sample processing For instructions on human milk preparation and handling see the Miris HMA User Manual Use a sample container which will give a sample depth of at least 2 cm see Figure 4 allowing some extra space for sample movement If using conical tubes a tube holder i necessary to keep the tube upright and sta...

Страница 11: ...4 After each use of the Miris Ultrasonic Processor wipe the probe with a tissue At the end of the day wipe the probe and all contaminated areas with a cloth dampened with Miris CLEANER Polish the probe tip with a piece of emery cloth Turn off the Miris Ultrasonic Processor by pressing O Figure 4 Sample container placement ...

Страница 12: ...a faint odour soapy chemical and no colour Materials provided Materials required but not provided Liquid concentrate in 50 ml container 1000 ml glass or plastic container 950 ml distilled or deionized water Do not use if the liquid is cloudy or precipitations are visible Use before Expiry date for unopened tubes is 1 year from production date Use solution within 3 months of preparation Environment...

Страница 13: ...g with fine grit emery cloth Polish off the eroded part only and no more or the probe will wear out quicker Wipe spillages immediately Clean the instrument surface and inside using a cloth dampened with Miris Cleaner If necessary a mild disinfectant can be used to clean the probe and the instrument surfaces Wipe off the probe after every milk sample using a tissue Clean the probe with a cloth damp...

Страница 14: ...nto an electrical outlet that provides a different voltage from that required The probe is not secured properly A fuse has failed Check the voltage on the back of the instrument Make sure the probe is properly attached Use two wrenches hand tightening is not sufficient Restart the instrument Contact Miris support miris se or your local distributor Foamy sample uneven homogeni sation The probe is i...

Страница 15: ...quency 20 kHz Display LCD Probe Titanium alloy Ti 6Al 4V Probe size Length 138 mm diameter Ø 3 mm Probe processing capacity 3 10 ml Standards 2014 30 EU Electromagnetic Compatibility Directive 2014 35 EU Low Voltage Directive 2011 65 EU RoHS 2 Directive UL CSA EN 61010 1 2013 3rd Edition created and disrupted within microseconds a phenomenon called cavitation When the air bubbles disrupt the relea...

Страница 16: ...www MirisSolutions com ...