Spirobank II
Rev.1.6
Page 33 of 35
User manual
•
If no beeps are heard while moving the device, proceed to clean the turbine
4.2
Cleaning of the device
Clean the device once a day or every time changes the patient. Use only the substances and methods listed in this chapter to clean the
device.
Recommended cleaning agents are:
•
Mild soap (diluted)
•
Sodium hypochlorite bleach (10% diluted)
•
Hydrogen peroxide (1.5%)
•
Alcoholic Solvents
Moisten a soft cloth with a recommended solution, but not so much that the cloth drips, and lightly wipe the surface for 30 seconds.
Let it air dry. Do not use ketonic solvents and aromatic solvents. Never put the device into water or other fluids.
4.3
Oximetry sensor cleaning
The reusable finger sensor must be cleaned every patient change, so clean the sensor before to use this on a new patient.
Clean the sensor with a soft cloth moistened with water or a mild soap solution. To disinfect the sensor, rub with isopropylic alcohol.
Allow the sensor to dry completely after cleaning.
Do not use any abrasive or caustic material to clean the sensor.
WARNING
Do not sterilize by irradiation, steam or by using ethylene oxide.
Unplug the sensor from the device before cleaning or disinfecting it.
The sensor included with the
Spirobank II
is made with latex free material.
4.4
Battery charging
Turn on
Spirobank II
and the following icon will appear on the main screen showing the charge level of the battery pack:
The maximum charge level is displayed with all 6 bars inside the battery.
If only one bar is shown or if the device will not even turn on the battery pack must be recharged in the following manner:
•
Plug the battery charger into a socket and the battery charger cable into the micro
USB connector of the device; the device in this phase is always turned on
•
When the charging is complete the battery icon will display all six bars.
•
At this point disconnect the battery charger from the device.
SpO2
WARNING
It is recommended not to use the device while the battery is charging.
Always disconnect the battery charger from the device when the charge cycle has terminated.
WARNING
Operator shall not touch simultaneously the patient and the parts of non-medical equipment that are accessible to the operator
during routine maintenance after removal of covers without the use of a tool.