Table 18. Status word values description
Status
word
Priority
level
12
Description
Recommendations
00
12 (lowest)
Normal operating mode, temperature
is static.
–
10
2
Sensor is warming up.
Do not perform span calibration.
11
3
Data request rate is more than 1 Hz.
Do not perform span calibration or
zeroing. Decrease request frequency.
21
10
Dynamic
temperature
mode
(temperature changes faster than by
0.6 °C/min).
Do not perform span calibration or
zeroing.
22
9
Dynamic
temperature
mode
(temperature changes faster than by 2
°C/min).
Do not perform span calibration or
zeroing.
24
7
Dynamic temperature mode; zero
shifts to negative value.
Do not perform span calibration or
zeroing.
30
4
One of the signal values (U
s
or U
ref
) is
lower than it is allowed
Possibility of condensed moisture on
optical elements. Dry sensor. After
drying, if this status lasts longer than
for 20 minutes at stable conditions,
replace sensor.
31
8
Zero shifts to negative value.
Possibility of condensed moisture on
optical elements. Dry sensor. After
drying, if this status lasts longer than
for 20 minutes at stable conditions,
perform zeroing (see Appendix D.1 for
procedure description).
40
6
Exceeding temperature limits.
Check the ambient temperature.
50
11
An abrupt signal change (due to
feeding a gas mixture to sensor
atmosphere) or increased noise on
sensor optical elements.
Do not perform span calibration or
zeroing.
If this status lasts longer than for 20
minutes at stable conditions, replace
sensor.
51
5
Complex status. Technological failure. Contact support for details
90
1 (highest)
Firmware
failure
(flash
memory
issues).
Contact support for details
1XX
–
Sensor
operates
in
low
power
consumption mode.
–
NOTE: If any status word value
except “00” and “21” appears, metrological properties of sensor
are not provided.
12
Sensor condition is characterized by set of status word parameters. <F> command returns only one
status word value in accordance with priority. To view all parameter values use command <DATAE2>.