96
MiniCAT™ User Manual v5.2
© 2014 Xoran Technologies LLC
Leakage Radiation
Less than 100 mR/Hr
Leakage Technique Factors
(SourceRay Source)
125 kV, 7 mA, 0.014 s pulse width, 30Hz, 8% duty cycle
Leakage Technique Factors (ISI
Source)
125 kV, 7 mA, 0.018 s pulse width, 15Hz, 16% duty
cycle
Beam Quality - half-value layer
(HVL) (SourceRay Source)
9.4 mm equivalent Al @ 120 kVp
9.7 mm equivalent Al @ 125 kVp
Beam Quality - half-value layer
(HVL) (ISI Source)
10.6 mm equivalent Al @ 120 kVp
10.9 mm equivalent Al @ 125 kVp
Minimum Permanent Filtration
(SourceRay Source)
11.7 mm Al equivalent
Minimum Permanent Filtration
(ISI Source)
10.6 mm Al equivalent
Measurement Criteria for Peak
Voltage, Current, and Pulse
Width
50% - 90 % of the voltage waveform, 125 kVp, 58.8
mAs
Содержание xoran
Страница 1: ...USER MANUAL ...
Страница 86: ...83 MiniCAT User Manual v5 2 2014 Xoran Technologies LLC FRONT VIEW TOP VIEW ...
Страница 97: ...94 MiniCAT User Manual v5 2 2014 Xoran Technologies LLC Sinus 10s 150 T Bone 20s 600 T Bone 40s 600 ...